NEGATIVE DISCHARGE TIME MEASUREMENTS

To provide a NEGATIVE voltage output, touch the plate

located underneath the Charger, and press the switch

downward at the same time.  Momentarily touch the

Charger’s output terminal to the conductive plate 

attached to the Field Meter.  The meter reads 

approximately -1.10 kV.  By using a stop watch or other 

timing device, determine the time needed for the 

voltages to decrease from -1.10 kV to -0.10 kV.  This is 

the negative discharge time.

IMPORTANT:

 A ground path must be provided between 

the touch plate of the Charger and the ground reference 

of the Field Meter.  This is normally provided by holding 

the Charger in one hand and the Field Meter with 

Conductive Plate in the other.

The battery in the Charger should be replaced annually 

or when it is unable to provide approximately ±1100V.

CLEANING

It is important to keep the insulators on the adapter plate

clean and free of contaminates that may cause surface

leakage.  To test the performance of the adapter plate, 

charge the plate and note the discharge rate in a nonion

-

ized area.  The self discharge rate to 10% of original

voltage should not be less than five minutes.
The area around the aperture of the Digital Static Field

Meter must be kept clean to ensure accurate, drift-free

readings.  Never touch the aperture with anything.  To

remove dust or other particulate matter, use low-

pressure instrument-grade air.  To remove more severe 

contamination, spray or flush with the smallest practical 

amount of clean technical-grade of isopropyl alcohol.  

Then allow the instrument to air dry for several hours.

Specifications

DIGITAL STATIC FIELD METER
Measurement Range (switch selectable)

Low Range: 

 

0 to ±1.99 kV / inch

High Range: 

 

0 to ±19.99 kV / inch

Measurement Accuracy

Voltage Monitor Output: > ±5% of reading ±10 mV

Voltage Display:  

> ±5% of reading ±2 counts

Measurement Stability

±10 counts

Automatic Shutoff

Unit will shut off after 20 minutes after last switch activity

Power Requirements

One (1) 9V alkaline battery

Operating Time

Greater than 50 hours, with new battery at 21°C 

continuous usage

Operating Conditions

Temperature:   

10-30°C

Relative Humidity: 

Up to 80%, non-condensing

Altitude: 

 

Up to 2,000 meters

Dimensions

0.94" H x 2.75" W x 4.94" L

(23.9mm H x 69.9mm W x 125.5mm L)

Weight (with battery)

4.9 oz

(153 g)

Voltage Monitor Connection

2.5mm audio jack

CE Certified

Figure 8. Taking decay measurements

Maintenance

The Digital Static Field Meter is factory calibrated and no

maintenance is required. If for any reason you believe 

the Meter is not working correctly, please contact Desco

Customer Service.  CAUTION - There are no user

serviceable parts. Any unauthorized service will void the 

warranty and result in additional repair charges.

Note:

 This Meter is a precision instrument and should 

not be subjected to dropping as that would void the 

warranty.

BATTERY REPLACEMENT

The Digital Static Field Meter operates from a standard 

9 VDC alkaline battery.  Battery life is in excess of 50 

hours under normal use.  When the battery voltage 

drops below 6.5 V, “BAT” will appear on the display.  To 

change the battery, slide the battery cover down at the 

back of the Meter and remove the battery from the 

battery clip.  Replace the battery with a fresh one and 

reinstall the battery cover.  The battery should be 

removed from the Meter if its is to be stored for an 

extended period of time.

Содержание 19493

Страница 1: ...Charged insulators in the ESD protected area can adversely impact quality productivity and reliability When any object becomes electrostatically charged there is an electrostatic field associated with that charge If an ESDS ESD sensitive device is placed in that electrostatic field a voltage may be induced on the device If the device is then momentarily grounded a transfer of charge from the devic...

Страница 2: ...K VIEW E F A HOLD Button Press to freeze the reading on the display Press again to return to normal measurement operation B RANGE ZERO Button Press to select the measurement range Press and hold to zero the Meter C Analog Output Jack A low voltage replica of the measured voltage is provided at this output The voltage at this output is 1 1000th 2 kV range or 1 10 000 20 kV range of the measured vol...

Страница 3: ...p connection is utilized to achieve accurate measurements INSTALLING THE ISOLATED PLATE ASSEMBLY The Digital Static Field Meter s case has two slots along its sides The top slot is closest to the face of the instrument Slide down the tabs of the Conductive Plate plate into the top slot of the Meter s case as far as they go see Figure 4 BATTERY CHECK The battery should be replaced when BAT is indic...

Страница 4: ...PERATING THE CHARGER The Charger has a momentary rocker switch that powers the unit Holding the switch forward backward supplies power to the output terminals POLARITY SELECTION The top of the rocker switch is labeled and the bottom is labeled To provide a POSITIVE voltage output touch the plate located underneath the charger and press the switch forward at the same time To provide a NEGATIVE volt...

Страница 5: ...h the smallest practical amount of clean technical grade of isopropyl alcohol Then allow the instrument to air dry for several hours Specifications DIGITAL STATIC FIELD METER Measurement Range switch selectable Low Range 0 to 1 99 kV inch High Range 0 to 19 99 kV inch Measurement Accuracy Voltage Monitor Output 5 of reading 10 mV Voltage Display 5 of reading 2 counts Measurement Stability 10 count...

Страница 6: ...ur Customer Service Department at 909 627 8178 Chino CA or 781 821 8370 Canton MA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY A...

Отзывы: