- 16 -
y
If calibration is completed, PASS or FAIL appears on the
primary display.
y
If both calibrations are passed, press ”CAL” key again to
store the calibration value to EEPROM.(Electrically Erasable
Programmable Read-Only Memory).
Display the measurement reading
Press “
CAL
” key for 2sec.
Enter to open calibration mode
Press “
CAL
” key
Start the open calibration with
30-seconds countdown
OPEN calibration is completed.
Pass or Fail will appear on the
primary display
* If FAIL, run the procedure from
the first step again.
Press “
CAL
” key
- 17 -
Enter short calibration mode.
* Short alligator test leads.
Press “
CAL
” key
Start the short calibration with
30-seconds countdown
SHORT calibration is completed
PASS
or
FAIL
will appear on the
primary display
* If FAIL, run the procedure from
the first step again.
Press “
CAL
” key
Calibration is completed
Return to testing mode.
When open or short calibration
fails. “
FAIL
” will appear on the
primary display.
y
If fails in calibration, please double check the DUT( device
under test ) is in the right calibration position and resume the
open and short calibration procedures.