SECTiON 18:

G137 Cavitron

®

 JET Plus Troubleshooting and Analysis

, cont 

This troubleshooting section is meant for use by qualified Cavitron

®

 Service Technicians.

SymPTOmS

CAuSES

CORRECTivE mEASuRES

No water flow from handpiece 

with no insert installed.

1. High dental office water pressure. 1. Install a water pressure regulator on the main  

    water supply line and reduce the pressure to  

    20-40 psi.

Water spray from insert is not 

properly covering the operating 

area of the activated tip.

1. Improper water flow adjustment.

2. P-style insert water tube  

    incorrectly aimed.

3. Insert or airpolishing nozzle   

    insert is partially clogged.

1. Refer to the Cavitron

®

 JET Plus Directions for  

    Use for instructions on water flow adjustment.

2. Use small smooth pliers, reposition the water tube  

    and direct the spray at the back of the insert tip.

3. Replace the insert or airpolishing nozzle insert.

Water drips from the handpiece 

when not operating.

1. Water solenoid valve leaking due  

    to trapped debris.

1. Try plugging the water supply hose into an air  

    source to blow out the dirt.  If the leak persists,  

    replace the Water Regulator/Solenoid assembly.   

    Be sure the external hose filter is installed.

Water leak from the handpiece 

while in operation.

1. O-ring worn on insert.

2. Water leak in plastic water line  

    at handpiece or inside the  

    JET-Mate™ Handpiece.

1. Replace the O-ring with genuine Cavitron

®

 O-rings.   

    O-rings are available in packs of 12:   

    Green O-Rings P/N 62605  

    Black O-Rings P/N 62351

2. a.   Unplug the JET-Mate™ handpiece from the  

          cable and replace the small O-ring on the  

          connector.  Part No. 79357 (12-Pack)

    b.   Replace the JET-Mate™ handpiece and/or  

          cable assembly.

Water flow not controllable 

by turning the handpiece flow 

control knob.

1. Malfunction of Water  

    Regulator/Solenoid assembly.

1. Replace the Water Regulator/Solenoid valve  

    assembly.  Adjust the water regulator to  

    specifications.

Intermittent activation or no 

activation when stepping on  

the Foot Pedal.

1. Foot Pedal batteries are weak.

1. a.   Check Foot Pedal battery condition.  

          Replace batteries as required.

    b.   Connect the auxiliary Foot Pedal cable  

          between the Foot Pedal and unit.  The unit  

          can be operated with the cable until the  

          batteries are replaced.

Содержание Cavitron JET Plus

Страница 1: ...i Cavitron JET Plus Ultrasonic Scaler Air Polishing Prophylaxis System Installation and Service Manual with Please read carefully and completely before operating unit...

Страница 2: ......

Страница 3: ...1...

Страница 4: ...2...

Страница 5: ...E Illuminates when the system is not functioning properly This display has three distinct modes Slow blink 1 blink per second means the system is not operating within factory specifications Fast blink...

Страница 6: ...ron 30K Ultrasonic Inserts 14 8 5 Cavitron JET Air Polishing Inserts 15 8 6 Tap OnTM Wireless Foot Pedal Operation 15 8 7 Accessories and User Replaceable Parts 15 8 7 1 Accessories 15 8 7 2 User Repl...

Страница 7: ...JET air polishing insert tip that polishes the tooth enamel without contact so there is less abrasion to enamel and no physical pressure or heat build up that could cause discomfort in sensitive patie...

Страница 8: ...f pacemakers and other medically implanted devices on the market Clinicians should contact the device manufacturer or the patient s physician for specific recommendations This unit complies with IEC 6...

Страница 9: ...shield gloves and protective gown Ultrasonics The Cavitron JET Plus unit works with Cavitron inserts as a system and was designed and tested to deliver maximum performance for all currently available...

Страница 10: ...ceable filter is supplied with your system See Section 10 System Care for replacement instructions Incoming water supply line pressure to the system must be 20 psi 138 kPa to 40 psi 275 kPa If your de...

Страница 11: ...est on a level surface Be sure unit is stable and resting on four feet Placing unit in direct sunlight may discolor plastic housing The system has been equipped with a Cavitron Tap OnTM Wireless Foot...

Страница 12: ...Foot Pedal over and using a Philips screwdriver carefully remove battery cover screw and battery cover If applicable remove used batteries and install two new AA batteries as shown Do not depress Tap...

Страница 13: ...the Diagnostic Display The graphics will begin to blink in a sequential pattern representing the synchronization mode This mode will last 5 to 6 seconds 6 During this mode press the red button located...

Страница 14: ...ultrasonic power level control knob fully counterclockwise until a click is heard Rinse mode is for use during an ultrasonic scaling procedure when lavage is desired with minimal cavitation Diagnosti...

Страница 15: ...cator Illuminates when the Boost Mode has been activated with the Tap OnTM Foot Pedal Turbo Mode When pressed extra power is delivered to the system up to 25 This will increase the stroke of the tip T...

Страница 16: ...the insert port Powder Delivery Port Creates an airtight seal between the air polishing insert and the handpiece Replace when wear is noticed or powder is leaking at nozzle interface Insert Port The C...

Страница 17: ...eliminates the need to hold the pedal down Tapping the foot pedal once activates ultrasonic power or rinse mode for approximately 4 minutes Tapping the foot pedal while in Tap On TM mode disables the...

Страница 18: ...ressing the Turbo button on the scaling unit increases the unit s ultrasonic power up to 25 When you need more power for an extended period of time simply press the Turbo button on the display panel a...

Страница 19: ...sensitive tooth to another and then return 9 7 Air Polishing Powder Bowl Use only Cavitron Prophy Powders in your Cavitron JET Plus Combination System Any other substance or additives may clog the sy...

Страница 20: ...on High Volume Evacuator with the aid of a dental assistant is recommended When performing air polishing without the aid of a dental assistant the use of a saliva ejector and or aerosol reduction devi...

Страница 21: ...the Power Level Control and Lavage Control to your preferred operating position for ultrasonic scaling For air polishing place a sterilized JET Air Polishing Insert into the handpiece and adjust the P...

Страница 22: ...scolored the filter should be replaced to prevent reduced water flow to the Cavitron JET Plus system A 10 pack of replacement filters is available by ordering Part Number 90158 from your local DENTSPL...

Страница 23: ...OnTM Foot Pedal to the first position The system should dispense water If none of the Tap OnTM Foot Pedals operate the system continue to the next step 2 Resynchronize one Tap OnTM Foot Pedal to the s...

Страница 24: ...ow powder level or empty powder bowl Fill Powder bowl 2 Blocked JET Air Polishing insert nozzle Clean nozzle using supplied tool 3 Powder Cap loose Turn Main Power switch to the OFF O position Tighten...

Страница 25: ...hetic or oxygen According to medical device directive IIA rule 9 Section 15 Disposal of Unit U S Dispose of the system components in accordance with state and local laws EU Dispose of in accordance wi...

Страница 26: ...61000 3 2 Class A Not Applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not Applicable The Ultrasonic Scaler is suitable for use in all establishments including domestic establishments a...

Страница 27: ...ite surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the high...

Страница 28: ...ectromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distance between Portable and mobile RF communications equipment and the mo...

Страница 29: ...ove insert B Turn Main Power Switch OFF O position Wait five seconds Turn unit ON I position C Operate Purge function D If service indicator still blinks refer to Section 11 2 Technical Support and Re...

Страница 30: ...it 1 Provide adequate air circulation around unit Slo Blo Fuse F3 and or F4 failed 1 Short in Power supply assembly 2 Short in Power Drive PC Board assembly 1 Replace the Power Supply assembly 2 Repla...

Страница 31: ...aking 3 Replace JET Mate Handpiece 4 Replace JET Mate Handpiece 5 Install a working JET Mate Handpiece on the cable Unplug the Handpiece cable connector at J3 of the Power Drive PC Board and check the...

Страница 32: ...Cavitron insert and check operation 4 Loop handpiece cable around arm or support with finger to prevent water restriction 5 Replace with a new Cavitron insert Worn inserts require higher power settin...

Страница 33: ...ng the water supply hose into an air source to blow out the dirt If the leak persists replace the Water Regulator Solenoid assembly Be sure the external hose filter is installed Water leak from the ha...

Страница 34: ...in the handpiece for the Boost mode to activate 3 Replace the Foot Pedal Follow the Cavitron JET Plus Directions for Use and Service Manual instructions for Foot Pedal synchronization Scaling inserts...

Страница 35: ...the Cavitron JET Plus unit to DENTSPLY for factory certified service 4 Return the Cavitron JET Plus unit to DENTSPLY for factory certified service Info Center Service light stays lit 1 Unit is install...

Страница 36: ...to the H or 12 o clock position 2 Dump out the powder and clean out the bowl assembly 3 Remove the tube attached to the bowl bottom L Nozzle and check airflow 4 Use the cleaning tool and clean out th...

Страница 37: ...the Cavitron JET Plus unit to DENTSPLY for factory certified service Cavitron JET Plus unit has no bleed air 1 The Air Polishing insert is clogged 2 The Cavitron JET Plus unit air supply is not hooke...

Страница 38: ...to specifications 3 Clean out the Y Connector with a cleaning wire Powder bowl cap cannot be unscrewed 1 Cavitron JET Plus unit is still powdered up 2 Powder bowl chamber is not fully depressurized 3...

Страница 39: ...moving the protective cover on the power entry module 2 Unplug the power cord from the rear Power Entry module 3 Follow the steps above in Section A for Top Cover Removal 4 Ensure that the Power Switc...

Страница 40: ...or 8 Cut the cable tie for the Snap On Ferrite block on the Water Regulator Solenoid The latch to open the Snap On Ferrite block is positioned downward Rotate the block to gain access to the latch Ope...

Страница 41: ...to the Bottom Housing using the 6 x 1 2 Hi Lo screws Place the ground terminal and lockwasher under the left screw 12 Connect the small clear Handpiece tube to the top port of the Water Regulator Sol...

Страница 42: ...arrier from the unit Pull the barrier away from the heatsink to release the two sided tape and lift it out of the unit 4 Unplug the BLK RED wire connector from the Power Drive PC board at J1 5 Unplug...

Страница 43: ...the cable 7 Calibrate and final test the Cavitron JET Plus unit using DENTSPLY Professional Product Service Standard Operating Procedure PS 00176 8 Align the Top Cover on the Bottom Housing and secure...

Страница 44: ...ver to remove the two 6 32 x 5 16 screws under the Air Manifold Assembly Position the replacement Air Manifold assembly over the same mounting bracket holes and fasten with the removed Phillips screws...

Страница 45: ...ack under the molded tab located adjacent to the water inlet hole Install the two 6 x 1 2 Hi Lo screws at the front corners of the Air Water chassis assembly The Ground wire terminal must be reinstall...

Страница 46: ...693 2 18 0 125 1 8 Yellow Tube P N 625036012 3 17 75 25 1 8 Red Tube P N 61631 4 0 50 125 1 8 Gray Tube P N 625036010 5 5 50 25 1 8 Gray Tube P N 625036010 6 0 50 125 1 8 Clear Tube P N 61693 7 12 25...

Страница 47: ...45 Section 20 Air and Water Flow Diagram G 137 Cavitron JET Plus with Tap On Technology Air and Water Flow Diagram...

Страница 48: ...p In 9 4 592051001 Rubber Feet Self Adhesive 10 2 592055001 D Section Seal 1 Inch 11 1 568270002 Power Entry Module 12 4 586172005 6 19 x 1 2 HI LO Torx Pan Head Screw 1 745 06002 6 Lockwasher For Gro...

Страница 49: ...586174003 6 19 x 3 8 HI LO PPH Screws 6 1 61679 Straight Fitting 1 16 Barb x 10 32 7 1 60386 Washer Buna Fibre 8 1 81730 Pinch Plate 9 1 26109 Dynamic Seal 10 1 26105 Piston 11 1 80688 Pinch Cylinder...

Страница 50: ...ated 5 1 81881 Molded Overlay Label Info Center 6 14 586172002 4 24 x 1 2 Hi Lo Screws 7 6 62970 Washers 8 1 81852 Info Center PC Board 1 81720 Protective Cover Info PC Board 1 592059001 Foam Seal For...

Страница 51: ...Skirt 6 2 81710 L Shape Brackets 7 2 60898 6 32 x 3 8 Machine Screws 8 1 81712 Powder Bowl Housing 9 2 586172001 4 24 x 5 16 HI LO Torx PH Screw 10 1 60382 Filter Element 11 1 60384 O Ring 12 1 623071...

Страница 52: ...50 Service Parts continued G 137 Cavitron JET Plus Power Supply ITEM QUANTITY PART NO DESCRIPTION 1 1 81661 1 24 V Power Supply...

Страница 53: ...51 Service Parts continued G 137 Cavitron JET Plus Power Drive PCB Bracket Assembly ITEM QUANTITY PART NO DESCRIPTION 1 1 81651 Power Drive PC Board Bracket Assembly...

Страница 54: ...52 Service Parts continued G 137 Cavitron JET Plus Info Center PC Board Assembly ITEM QUANTITY PART NO DESCRIPTION 1 1 81852 Info Center PC Board Assembly 1 81720 Cover Protective...

Страница 55: ...53 Service Parts continued G 137 Cavitron JET Plus Main Controller PC Board Assembly ITEM QUANTITY PART NO DESCRIPTION 1 1 81848 Main Controller PC Board Assembly...

Страница 56: ...Tap On Foot Pedal ITEM QUANTITY PART NO DESCRIPTION 1 1 81872 Foot Pedal Tap On Packed w Synchronizing Instructions and 4 AA Batteries 1 81680 Battery Door and Screw Kit 1 81681 Foot Pedal Pad Bottom...

Страница 57: ...Batteries 1 81663 2 Auxiliary Foot Pedal Cable 1 90088 Air Filter with Fittings 1 81721 Air Filter Hose Assembly 1 90125 Water Filter Hose Assembly 90158 Water Filter 10 Pack 2 61693 Clear Tubing per...

Страница 58: ......

Страница 59: ......

Страница 60: ...nce Australia Phone 7531 583 0 Phone 1 30 14 77 77 Phone 61 3 9538 8280 U nited Kin gd o m Italia Canada DENTSPLY Ltd DENTSPLY DeTrey Italia s r l DENTSPLY Canada Building 1 Via A Cavaglieri 26 161 Vi...

Отзывы: