Cleaning Instructions

Alternate current

Power On 

 

I

Power Off  

O

Caution, Hot Surface

9

10

PRESS HERE TO RAISE

OBTURATOR HOLDER

PRESS HERE TO LOWER

OBTURATOR HOLDER

PLACE OBTURATOR

HERE

HEATING

CHAMBER

HEATING

CHAMBER

GREEN LIGHT TO

INDICATE OVEN HAS

BEEN TURNED ON

PRESS HERE FOR

OBTURATOR SIZES .04 -.08

PRESS HERE FOR

OBTURATOR SIZES .10 -.12

PLACE OBTURATOR

HERE

AFTER SELECTING

PROPER SIZE BUTTON,

PRESS “START LEFT”

FOR LEFT OBTURATOR

HEATER

AFTER SELECTING

PROPER SIZE BUTTON,

PRESS “START RIGHT”

FOR RIGHT OBTURATOR

HEATER

POWER ON/OFF

SWITCH

4.  Push the holder down (arrow down) until you hear it click.
5.  Push the button which corresponds to the size of the GT Obturator 

you wish to heat. Then push the “start left” button. The obturator is 
now being heated. The heating time varies depending on obturator 
size and is regulated automatically. 
The heating times for each button are as follows:

 

Size 

Signal Time 

Tolerance

 

.04 -.08  20 seconds 

± 5 seconds

 

.10 -.12  41 seconds 

± 7 seconds 

6.  After the first signal “beep,” the obturator is ready for use. Push the 

obturator holder (arrow up) and guide the holder with your finger. 
Take the obturator carefully out of the holder by pulling it toward 
you, making sure not to scrape the obturator on any part of the 
holder.

7.   You may leave the obturator in the GT Obturator Oven for up to 90 

seconds after the first signal “beep.” It will keep the obturator at the 
appropriate temperature and ready for use. The oven will “beep” 
every 15 seconds to remind you that the obturator is still in the oven. 
After 90 seconds, the heating element will switch off automatically.

8.  As an extra safety precaution, after you’ve taken the obturator out 

of the holder, switch off the heating element by pushing the start 
left button again. Keep in mind that you won’t be able to switch the 
heating element off during the heating cycle.

9.  If you want to heat more obturators, you may alternate using the left 

and right holders to continue your work efficiently. Wait until after 
the signal “beeps” for the first holder, then you may immediately 
start to heat the other one. You will be unable to heat both sides at 
the same time.

10.  When you’re finished using the oven, use the power on/off switch 

located at the back of the oven to turn it off. Then unplug the oven 
until you are ready to use it again.

Cleaning Instructions

Clean oven after each use. Any gutta-percha remaining on the holder 

or in the heating chamber may be effectively removed with Orange Solvent 
(generic dental solvent) or any cotton pledget.

Technical Specifications

Voltage rating: 110 VAC, 50/60 Hertz, range of the supply v/- 

10%, Fuse 2 Amps Slow. 

Содержание ProSystemGT

Страница 1: ...Complete Endodontic System Directions For Use For the removal of dentin shaping and obturating of the root canal...

Страница 2: ...he canal 5 Clean file flutes frequently during instrumentation after the file is removed from the canal 2 6 Frequently irrigate and lubricate the canal throughout the procedure 7 Exercise caution in t...

Страница 3: ...regardless of canal form and size 3 Procedures for Using the Aseptico Endo DTC Motor Preset torque settings are recommended for each file series They are given without guarantee of risk or fracture To...

Страница 4: ...ubmerge the obturator in the heating 5 In the presence of NaOCl start crown down shaping with the 20 10 GT file and progress to smaller tapers if necessary crown down until one of these 20 Series GT f...

Страница 5: ...with air or GT absorbent points See GT Fiber Post directions for use for steps to appropriately place the post Post Drill Sterilization Drills must be cleaned and sterilized before every use Using dis...

Страница 6: ...obturator on any part of the holder 7 You may leave the obturator in the GT Obturator Oven for up to 90 seconds after the first signal beep It will keep the obturator at the appropriate temperature a...

Страница 7: ...nson City TN 37604 1 800 662 1202 1 800 597 2779 fax www tulsadentalspecialties com GT instruments and obturators made in USA 2009 DENTSPLY Tulsa ental pecialties IM32 2 09J For More Information Or To...

Отзывы: