Page 7 of 9

AEU-20T Manual

STERILIZATION & MAINTENANCE

CONSOLE

-  The exterior of the console may be

cleaned by wiping with a soft cloth moistened with mild
detergent or disinfecting solution.

HANDPIECES

-  Thorough cleaning and lubrication of

handpieces after each use and before sterilization is
very important to ensure proper operation and service
life of the handpiece.  Follow the instructions provided
with the handpiece purchased for complete
maintenance instructions.

MOTORS

-  The entire motor assembly supplied with

your unit may be sterilized by autoclave, chemclave, or
ethylene oxide gas processes.  Follow the sterilizer
manufacturer's recommended procedure.

Before autoclaving the motor and cord assembly the
‘E’ type coupling on the motor must be plugged.  A
plug with sealing o-ring is supplied with the motor.
Remove the handpiece from the motor and install plug
as shown at lower right.

The included motor/handpiece holding cradle is also
fully autoclavable.

Always install the autoclaving plug into motor
and motor connector before autoclaving.

If the autoclaving plug for the motor becomes
difficult to remove or install, apply silicone
grease or similar lubricant to the o-rings.

Loosely coil the motor cord when autoclaving.
Avoid crowding the motor when autoclaving.

Turn black connector ring only.

CAUTION FOR ALL STERILIZATION:

6

Do not exceed 133° C or 275° F.

6

Do not submerge in any solutions.

6

Do not use ultrasonic cleaners.

WARNING

DO NOT

attempt to disassemble

the motor or motor connector.

DO NOT

oil or lubricate the motor.

DO NOT

attach a handpiece to the

motor while the motor is running.

DO NOT

bend motor cord sharply.

Failure to comply with any of the above

instructions may void your warranty.