5 Operation
5.5 Main home screen
Dentsply Sirona
Operating Instructions SiroLaser Blue
42
66 56 842 D3648
D3648.201.03.02.02 10.2018
USB port
!
USB
To make sure that the USB port is available, it is checked within the self
test.
If the appropriate symbol for USB port error is not displayed on the
home screen, the USB port works properly.
If the symbol is displayed on the home screen, the USB port is
defective.
NOTE
If the USB port error is displayed on the home screen as defective,
please contact your local dental dealer or an authorized Customer
Service Department for technical support.
The SiroLaser Blue will remain functional but the download of the
history file / software updates are not possible.
Calibration check
Cal. 30d
Dentsply Sirona recommends a calibration check with external
powermeter to be performed every twelve months, see chapter
"Calibration check using an external power meter [→ 53]".
The information for 'next calibration check' is only displayed on the
home screen for the first time after restart or log-in (in month).
For the last 30 days it is displayed continuously on the home screen.
After excess of the service interval the days are displayed with a minus
[-] and red coloured.
The laser remains fully functional during this time.
Time to service
30d
The safety test is a mandatory test for all medical devices. The
SiroLaser Blue needs to be tested once every two years.
The information for 'time to service' is only displayed on the home
screen for the first time after restart or log-in (in month).
For the last 30 days it is displayed continuously on the home screen.
After excess of the service interval the days are displayed with a minus
[-] and red coloured. The laser remains fully functional.
NOTE
Legal regulations require a regular safety test of the performance of a
laser device. The SiroLaser Blue needs to be tested once every two
years. Please contact your local dental dealer or an authorized
Customer Service Department for technical support.
If internal or external calibration check failed after last calibration a
warning screen pops up after self test. The laser remains fully
functional.
NOTE
Please contact your local dental dealer or an authorized Customer
Service Department for technical support.