Reprocessing Instructions

Important:

 Use only products recommended by Dentsply 

Sirona to routinely clean and lubricate your Nupro RDH™ 
Handpiece. 

There are two Maintenance procedures for your Nupro RDH™ 
handpiece: (1) for maintenance using the Midwest Plus™ 
Maintenance Kit; (2) for maintenance using Midwest Plus™ 
Aerosol Spray. Follow the appropriate maintenance procedure 
below before using your new RDH handpiece.

This procedure recommends the steps to follow between 
patient use to prevent cross-contamination and to ensure 
continued trouble-free operation of the handpiece:

5.  Apply one drop of Midwest Plus Handpiece Lubricant on 

gears inside head.  DO NOT fill head full, as excess lubricant 
will leak from seal.

Drive
Air
Tube
(Gold)

1. Maintenance Using  
  Midwest Plus™ Maintenance Kit

Clean External Surface

1.  Remove prophy cup by holding sheath of handpiece and 

rotating cup in counterclockwise direction.

2.  Disconnect handpiece from its supply hose. Spray exterior 

surface of handpiece with Midwest Plus™ Handpiece Cleaner.

3.  Scrub and rinse external surfaces of handpiece under warm 

tap water.

4.  Thoroughly dry using gauze, paper towel or air syringe.

Lubricate and Expel

1.  Place three drops of Midwest Plus Handpiece Lubricant into 

drive air tube (above).

2.  Run handpiece on Midwest Handpiece Air Station or other 

air source until it reaches normal speed. This ensures that 
the internal working parts of the motor are completely 
lubricated. Wipe off any excess lubricant on outside of 
handpiece with dry 2x2 or tissue. 

3.  Proceed to Sterilization.

2. Maintenance Using  
  Midwest Plus™ Aerosol Spray

Scrub and Rinse

1.  Remove prophy cup by holding sheath of handpiece and 

rotating cup in counterclockwise direction.

2.  Disconnect handpiece from its supply hose. 

3.  Scrub and rinse external surfaces of handpiece under warm 

tap water.  Use sponge to scrub. 

4.  Dry thoroughly using gauze, paper towel or air syringe.

Clean • Lubricate • Expel

1.   Clean and lubricate with one quick spray of Midwest Plus™ 

Aerosol Spray using the Midwest Backend Attachment 
nozzle into the drive air tube of the handpiece. The drive air 
tube is the shorter of the two larger tubes and is gold on 
many handpieces. 

2.   Run the handpiece on Midwest Handpiece Air Station or 

other air source for 30 seconds or until normal speed is 
reached, and all excess spray is expelled. Dry thoroughly.

3.  Proceed to Sterilization.

Sterilization

Enclose the handpiece in a sterilization pouch intended for 
steam sterilization. Place the pouched handpiece into the steam 
sterilizer per the manufacturer’s instructions.

Gravity Steam Sterilization

 

Full cycle: 121°C (250°F) for 30 minutes

Full cycle: 132 °C (270°F) for 15 minutes

Dynamic Air Removal

 

Full cycle: 132°C (270°F) for 4 minutes

Full Cycle: 135°C (275°F) for 3 minutes

Keep handpiece in sterilization pouch until ready to use in the
operatory. No further lubrication is required before use.

Other Maintenance

Weekly or as needed: Place two drops of lubricant on swivel 
junction (as shown). Twist swivel to work lubricant into  
junction. Wipe handpiece clean.

6.  Apply one drop of Midwest Plus Handpiece Lubricant to 

gear teeth on handpiece sheath.

7.  Reassemble and continue with “Sterilization” in the  

Maintenance procedure.

Repair

Should your Nupro RDH™ Handpiece need repair, call  
1-800-800-7202 to schedule a free pick-up.

Send to:

Dentsply Sirona
Attn: Air Repair
500 W. College Ave
York, PA 17401

Provide the following information for your repair:

 

Your complete name and address

 

Dealer name and address

 

Nature of repair

Air Repair offers:

•  Repairs DOOR TO DOOR IN 2 BUSINESS DAYS

•  Genuine Parts

•  Expert Troubleshooting and Diagnosis

•  Repair Service Warranty equal to the new product warranty

Parts  

 

 REF

Midwest Plus

 

8 oz. Spray Cleaner 

380145 

Maintenance Kit

 

& 2 oz. Lubricant Dropper

Midwest Plus

 

8 oz. Spray Bottle 

380140

Handpiece Cleaner

 

 

Midwest Plus

 

2 oz. Dropper 

380130 

 

Handpiece Lubricant

Midwest Plus

 

Cleaner/Lubricant 

380080

Aerosol Spray

 

(does not include nozzle)

Nozzle for Midwest

 

Midwest Backend/ 

380086

Plus Aerosol Spray

 

Attachment

Midwest Handpiece

  

includes box of 12 

380800 

Air Station

 

FREE Spray Guards 

Midwest

 

Box of 100 

380150 

Spray Guards

 

Spray Tips for

 

pkg. of 6 

380986 

Midwest Plus 8 oz.

 

 

Spray Cleaner

 

Prophy Head Wrench

   

230611

Color Bands,

                   Orange   760011R       Red     760012R 

set of 6                            Yellow    760013R 

Violet    760014R

 

 Blue       760015R 

Green   760016R

The Nupro RDH™ Hygienist Handpiece is compatible with a 
wide range of disposable prophy angles and prophy cups/
brushes.

Please see our website at dentsplysirona.com or contact your 
local Dentsply Sirona Customer Support Center.

Warranty

Your Nupro RDH™ Handpiece is guaranteed to be free of 
defects in materials and workmanship for a period of two (2) 
years from its date of purchase — provided that the air motor 
has been operated and maintained as prescribed in these 
instructions (at the discretion of Dentsply Sirona) and has not 
been subjected to apparent misuse, abuse or accident. All 
questionable, defective, or loose parts must be returned to the 
dealer, with air motor and proof of purchase, transportation 
prepaid.

(FOR Nupro RDH™ WITH RIGHT ANGLE ATTACHMENT ONLY 
  REF 740020 / 740030)

Every 40 prophies, or once a week:
1.  Unscrew angle from handpiece.

2.  Place angle in beaker of Midwest Plus Handpiece Cleaner 

and ultrasonically clean for ten minutes.

3.  Scrub and rinse under warm tap water.

4.  Dry thoroughly.