Dentsply Sirona Nupro Freedom Скачать руководство пользователя страница 2

Do not rinse inner 

module under 

the sink. 

Running water may damage  

internal components

Sheath and inner module 

must be completely dry 

after reprocessing and 

before assembly.

Excess moisture may damage the unit.

Do not use spray products 

or excessive liquids on 

inner module.

Excessive liquids may penetrate the 

inner module, causing failure.

Follow these top tips to help maximize the performance and lifespan of your Nupro Freedom

®

 System.

Do not drop unit while 

cleaning and during 

prophy procedures.

Place the unit on a tray or the  

cradle when not in use.

 Use registered EPA 

hospital disinfectants, 

such as 

VoloWipes

.

Unapproved chemicals may harm  

the inner module.

Top opening on plastic 

barrier must clear the 

metal nose cone of the 

inner module before 

attaching the sheath.

Gently clean charging 

contacts, the LED display 

window and inner module.

Care should be taken in cleaning the 

unit’s intricate workmanship.

Use pads of fingers to 

operate power button. 

Avoid using fingernails 

and sharp objects.

Sharp objects may wear button down. 

CARING FOR YOUR NUPRO FREEDOM

® 