6 Postprocessing
Instructions for Use Midwest Tradition / Tradition Pro / Quiet-Air
34
67 27 155 D3673.201.10.02.02
6
Postprocessing
6.1
After each treatment session
NOTE! Condition immediately, or at the latest, one hour after
treatment.
NOTE! Use only Midwest Plus Spray.
ü
The bur instrument must not be moving.
ü
Wear appropriate protective clothing.
1.
Purge the water and air channels on the treatment center for 30
seconds.
2.
Remove the bur instrument with tweezers.
3.
Predisinfect the high-speed handpiece directly at the treatment
center.
4.
Unscrew the high-speed handpiece from the hose.