61 30 608 D3497
42
D3497
.
076.01.09
.
02 10.2018
10 Upgrading
Dentsply Sirona
10.1 Upgrading to Superspeed
Service Manual inFire HTC speed
5. Cut the cable tie of the 2 cables of the main switch.
6. NOTICE! Always pay attention to the cable markings here! Pull out
the 4 cable lugs.
7. Release the two screws and remove the thermocouple cables.
8. Unscrew the regulator board (C), as illustrated in the overview image
(6 TX10 screws) and remove the board (no longer required).
9. Enter the serial number of the device. The serial number is changed
via service code F08560. The serial number of the device can be
found on the sticker attached to the back panel.