4 Transport and commissioning
Operating Instructions CEREC SpeedFire
4.6 Commissioning
båÖäáëÜ
4.6.3
Positioning the door insulation
1. Insert the positioning pin into the hole of the furnace door.
IMPORTANT
An update process interrupted due to a power failure means that the
control panel will be displayed when the unit is turned on and will
disappear immediately.
➢ In this case, press the Start button (A) on the control panel and keep
this pressed down until a beep sounds on the furnace and the
furnace restarts.
ª
This process can take 2-3 minutes.
CAUTION
Handling and disposing of the insulating material
Abrasion of insulating material in the furnace door area may occur,
which is visible as white dust. These deposits should:
- be wiped off with a moist cloth or
- be extracted with a vacuum cleaner with a fine particle filter (e.g. HEPA
/ Dust category H),
- not be blown away with compressed air,
- not be blown away by the wind.
Direct skin contact may lead to skin irritation. Please therefore avoid
contact with the skin and eyes and wear gloves and safety goggles as
necessary. In the event of contact with the skin wash the dust of the
insulating material off with clear running water.
Always dispose of the insulating material in a closed container (see also
"Disposal [ → 54]").
CAUTION
The unit must not be operated without door insulation.