Operating Instructions CEREC AC Connect
10 PC Diagnostic Tool
10.3 Test procedure
64 32 087 D3492
D3492.201.07.24.02 07.2018
81
Test result: Fail
An error was found on the PC-specific hardware. Replacing the PC
component or the PC may be required.
An image restore is not advisable.
1.
Check to find out which components did not pass the test.
2.
Carry out the appropriate steps in the chapter entitled
Troubleshooting [→ 82] .
3.
Note the number of the failed test and the error code if available.
Attach this information to the returned PC when replacing a PC.
End
"Technician Diagnostics"
test
1.
Remove the PC diagnostic tool CD from the drive.
2.
Switch the PC off by briefly pressing the on/off key.
10.3.3
Sirona Windows Diagnostics
General
The
"Sirona Windows Diagnostics"
test loads the PC-specific
components simultaneously over a long period of time. This period of
time is determined by the user.
As a result of this load
● the temperature in the PC is significantly increased compared with
normal operation.
A temperature malfunction or any existing faults are thereby
detected.
● Thanks to the temporal, unlimited test phase, sporadically occurring
faults are more likely to be detected.
Tip: Only conduct the test if at least one of the
"Customer
Diagnostics"
or
"Technician Diagnostics"
tests has been
completed with a
"pass"
and without errors.
Performing the test
Tip: Allow the
"Sirona Windows Diagnostics"
test to run for at
least 1 hour. For optimum results, allow the test to run overnight.
ü
You have started the
"Sirona Windows Diagnostics"
test
and Windows starts in the test environment. The
"Sirona
Windows Diagnostics"
test starts automatically.