8 Maintenance
8.1 Care, cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions CEREC AC Connect
62
64 32 087 D3492
D3492.201.07.24.02 07.2018
8.1.5.9
Use of CEREC Omnicam disposable sleeves
The CEREC Omnicam can be operated with disposable plastic sleeves
in order to ensure maximum infection control. The sleeves are available
using order number 66 32 264.
NOTE
The disposable sleeves cannot be used in the following
circumstances:
- with the CEREC AF and CEREC AI system,
- when using color analysis,
- when using CEREC Ortho SW software.
8.1.5.9.1
Attaching the disposable sleeve to the mirror sleeve of CEREC
Omnicam
1.
Remove one disposable sleeve from its packaging. Slide the
disposable sleeve up to the stop on the CEREC Omnicam mirror
sleeve, so that it is securely positioned on the CEREC Omnicam.
Hold the CEREC Omnicam in other places, not just on the sleeve in
order to prevent it from falling.
2.
Position the CEREC Omnicam in the holder, so that it can warm up
for 15 minutes prior to the intraoral exposure. Make sure that the
holder has been disinfected.
3.
In the case of intraoral exposures with disposable sleeves, the
CEREC Omnicam should be positioned closer to teeth than without
sleeves. If the 3D preview indicates artifacts, check whether the
window of the disposable sleeve is clean and wipe it down with
alcohol. Following each exposure, remove the disposable sleeve
from the mirror sleeve of the CEREC Omnicam and dispose of it
according to standard procedures.
8.1.5.9.2
Wipe disinfection for the camera and mirror sleeve
NOTE
Do not spray the CEREC Omnicam or immerse it in cleaning agents
or disinfectants!
1.
Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve [→ 56]").
2.
Use a new absorbent cotton gauze which is moistened with one of
the agents listed in the "Cleaning agents and disinfectants [→ 52]"
section. First wipe the camera enclosure and then the mirror sleeve.
The camera enclosure and mirror sleeve are disinfected through
this step.
3.
Next dry the mirror sleeve using a lint-free cloth.