8 Maintenance
8.1 Care, cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions CEREC AC Connect
52
64 32 087 D3492
D3492.201.07.24.02 07.2018
8
Maintenance
WARNING
Danger of touching live parts
If the housing is damaged, there is a possibility of touching live parts
inside the unit. If the housing is damaged, the unit must be put and
left out of operation until it has been professionally repaired.
NOTE
Regular inspection
Some countries have legal regulations which require regular safety
inspections of electrical devices or systems by the operator.
Dentsply Sirona would like to draw your attention to the fact that a so-
called "retest" (repeat test) must be carried out for the CEREC AC
acquisition unit at least every three years. In addition, this retest also
must be performed following every repair or retrofit of components
such as the PC, the PC power supply, the CEREC Omnicam camera,
and the camera cable.
NOTE
Annual maintenance performed by trained technical personnel is
recommended.
8.1
Care, cleaning, disinfection, and sterilization
NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by
Dentsply Sirona!
8.1.1
Cleaning agents and disinfectants
Alpro
● Minuten Spray classic
● Minuten Wipes
● Plasti Sept
● Plasti Sept Wipes
Merz
● Pursept-A
Dürr
● FD 312
Содержание CEREC AC
Страница 91: ......