Dentmate LEDEX WL-090 Скачать руководство пользователя страница 16

The LEDEX

TM 

WL-090 Dental Curing Light is an electric device 

designed to meet the worldwide electrical safety standards, which 
includes U.S.A. and Europe, so it’s safe and effective for all dental 
applications. 
To avoid electric shock:
•  Do not attempt to open or alter the unit in any way. Only the 

service centers authorized by DENTMATE can open the unit 
housing and repair the device. 

•  Do not put any foreign objects into the housing of the unit.
•  Use only the LEDEX

TM 

WL-090 cradle when recharging this 

product.  Never attempt to use any other devices for recharging. 

•  Connect the power plug into a suitably grounded and approved 

outlet. When you use an extend cable, make sure the grounded 
line is not interrupted. 

•  Always unplug the charging dock before disinfecting. 
•  Never use the power supply if the cord has been damaged.

4.3. To avoid electric shock (shock hazard)

As it is the case with all high-performance lights, the high light 
intensity results in a certain heat development. Prolonged 
exposure near the pulp and soft tissues may result in irreversible 
or reversible damage. Therefore,this high-performance curing light 
must be operated by trained professionals.

4.4.

 

Heat development (burn hazard)

Use only original spare parts, particularly DENTMATE batteries 
and charging bases. Do not short circuit battery. Do not store at 
temperatures above 40 °C / 104 °F(or 60 °C / 140 °F for a short 
period). Always store batteries charged. The storage period must 

not exceed 6 months. It may explode if it’s disposed of in fire.

4.5. Battery

Note : At least 10 mm gap between soft tissues and optical fiber 

light guide rod.

13

Содержание LEDEX WL-090

Страница 1: ...Instructions for use...

Страница 2: ...to the manufacturing of this product We sincerely hope that it will give you many years of trouble free use Please read and understand all the instructions before using this equipment and save this ma...

Страница 3: ...3 Contraindications 11 4 Warnings 12 4 1 User 12 4 2 Ambient conditions 12 4 3 To avoid electric shock shock hazard 13 4 4 Heat development burn hazard 13 4 5 Battery 13 4 6 Accessories 14 4 7 Repair...

Страница 4: ...oduct packaging Serial number Catalogue number Manufacturer Date of manufacture Class II AC Adapter Consult the Instructions for use This shows the Type B applied part Do not dispose it with normal ho...

Страница 5: ...inferior acrylic that optimizes light conduction and minimizes loss of light from source to tip Therefore it ensures the highest possible intensity of light at the light guide tip Advanced and high e...

Страница 6: ...wer supply Input AC100 240 V 50 60 Hz output DC5 V 2 A Optical fiber light guide rod 11 8 mm Filter Anti glare shield Disposable light guide sleeves Disposable curing light sleeves Instructions for us...

Страница 7: ...uring mode which is last used a b c a b c a c b d The display allows different information required by the user As shown above it comprises different zones identified from top to bottom as follows A d...

Страница 8: ...ich should to be kept during maintenance to prevent the liquid products from damaging the LEDs Then and insert the sterilized light guide into the handpiece Ensure the light guide is properly inserted...

Страница 9: ...is charging the display of the handpiece will glow and show the sign of charging battery 2 6 Operating modes Each time disinfect surfaces of the curing light light probes and anti glare cones before...

Страница 10: ...2 15 RAMP 5 10 15 20 25 30 35 40 1000 mW cm2 10 STANDARD 5 10 15 20 25 30 35 40 1000 mW cm2 10 HIGH 2 4 6 8 10 1800 mW cm2 10 FAST ORTHO 3 4 5 repeat 10 times 1800 mW cm2 10 TURBO 2 3 4 5 2400 mW cm2...

Страница 11: ...e is placed directly over the material in order to be polymerized Extend the curing time accordingly to increase the distance between the light source and the material Recommended curing modes Mode Ap...

Страница 12: ...prox half and extend the irradiation time Adaptive Status TURBO 4 A A When LEDEXTM WL 090 is operated frequently for long periods of time the temperature may become too high so the Safety mode functio...

Страница 13: ...moisture may cause electrical short circuit or dangerous malfunction 1 The curing time interval should exceed 5 seconds for each cycle 2 Verify the radiometer sensor which can impact the accuracy of...

Страница 14: ...th Urticaria solaris or erythropoietic protoporphyria or those who currently have treatments with photosensitising pharmaceuticals 4 While holding the unit steadily press and release the ON OFF Button...

Страница 15: ...rby If this happens reduce the interference by reorienting and repositioning the device or screening off the immediate environment The electromagnetic radiation emitted from this device is below the r...

Страница 16: ...terrupted Always unplug the charging dock before disinfecting Never use the power supply if the cord has been damaged 4 3 To avoid electric shock shock hazard As it is the case with all high performan...

Страница 17: ...es 2401 0005 Using other accessories spare parts may lead to increased emission of electromagnetic interference or to reduced electromagnetic interference immunity 4 6 Accessories 4 7 Repairs and defe...

Страница 18: ...show overheat sign If the temperature rises up to a really level please wait a moment and then use it again The intensity is too low If the result of intensity test in Standard mode is under 700 mW cm...

Страница 19: ...ly reprocessed and must not be contaminated 9 Warranty DENTMATE TECHNOLOGY Co Ltd warrants the product to be free of manufacturing defects for a period of one year from the date of purchase this is de...

Страница 20: ...dimensions 115 max x H68 mm Cradle weight 140 g Equipment class AC Adapter Class II Safety IEC 60601 1 EMC Electro Magnetic Compliance IEC 60601 1 2 Protection from electric shock Type B applied part...

Страница 21: ...has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices This medical device manufa...

Страница 22: ...WL 090 uses RF energy only for internal functions Therefore this RF emission is extremely weak and there is little chance of it creating any kind of interference whatsoever with nearby electronic equi...

Страница 23: ...r quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT for 0 5 cycle 5 UT for...

Страница 24: ...requency rangeb Interference may occur in the vicinity of equipment marked with the following symbol Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not apply...

Страница 25: ...W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 kHz to 2 5 GHz d 1 2 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3...

Страница 26: ...y 24872 Taiwan TEL 886 2 8976 9226 FAX 886 2 8976 9236 WEBSITE www dentmate com tw EMAIL info dentmate com tw EU representative TOPDENTAL PRODUCTS LTD 12 Ryefield Way Silsden West Yorkshire BD20 0EF U...

Страница 27: ...1711 ES V 1...

Отзывы: