6. Disinfection, Sterilization and Storage
6.1 Disinfection(Main Body)
•
Before and after use, wipe with just a dry cloth or with a cloth using Enzyssept.
•
Please dilute the QUAT disinfectant and pour it into Dry Wipes.
6.2 Sterilization (AMPOULE CASE)
Sterilization condition
Operating Condition
•
Temperature : 132
℃
(270
℉
)
•
Pressure : 150 kPa (1.6 kgf/
㎠
)
•
pre vacuum: 8 min(Wrapped)
•
Dry : 30 min
Autoclave
•
The use of the following substances is not permitted.
- High concentrations of organic acids and inorganic acids, hydrocarbons chlorinated
•
Sterilization condition(Please refer to the sterilization method for each country.)
6.3 Storage
•
Place DENOPS on CRADLE to charge after use.
•
Please keep it away from places where there is humidity or water.
•
Do not store near firearm.
•
Store the product at room temperature and out of direct light.