Read all instructions before use.
INTENDED USE:
DentalVibe
®
is a vibrating retractor and is intended for use
by dental professionals in a dental office setting. It has been
carefully designed to offer your patients a more comfortable
dental injection experience that is both safe and effective.
It does not modify the tissue or impart energy and is not
intended for therapeutic, diagnostic or any other purpose.
• The use of the DentalVibe varies between 10-120 seconds per application.
• Contraindication for use: Do not use in an area where an abscess is present.
DO NOT OPEN THE DENTALVIBE OR
CHARGING BASE.
The internal circuit boards are manufactured
in a static-free environment. For service or
battery replacement, please
call 1-877-503-VIBE (8423) for an authorized Bing Innovations,
LLC Service Center closest to you.
Caring for your
DentalVibe
®
SUPPLIED PARTS:
DentalVibe hand piece,
Charging base with Power Supply and Comfort Tips.
CLEANING INSTRUCTIONS:
Wipe handle clean with
plastic safe surface disinfectant. Thoroughly wet the
handle and ensure that the surface remains visibly wet
for 1 minute at room temperature (68°F/20°C).
•
Do Not place DentalVibe or Comfort Tip in an autoclave.
•
Do Not immerse DentalVibe or Charger in water.
•
Remove the power cord plug from the electrical outlet
before cleaning the charger.
WARNING:
IMPORTANT:
8
9
Do not place or store the charger where it can fall or be pulled into a sink.
Do not place the hand held unit or charger in water or other liquid.
Do not reach for a charger that has fallen into water. Unplug immediately.
Use this type B applied product only for its intended use as described in the manual.
Do not use attachments or chargers not recommended by the manufacturer.
Position the charger so the power supply is easy to unplug.
Never insert any objects into any opening of the unit or charger base.
Never operate the product if it has a damaged cord or plug, not working properly,
has been dropped, damaged or submerged into water. If any of these occur, contact
a service center and return the product for examination and repair.
Periodically check the cord for damage. If cord is damaged, ship the charging unit
to a Bing Innovations Service Center. A damaged or non-functioning unit should
no longer be used.
Do not touch the power cord with wet hands as this can cause electric shock.
When unplugging, always hold the power plug instead of the cord.
Do not put metal objects such as coins or clips on the charger. This may generate
heat, damaging the unit and heating the metal item presenting potential burn
hazards.
No User Serviceable Parts. Do not modify this equipment without authorization
of the manufacturer. If equipment is modified, appropriate inspection and testing
must be conducted to ensure continued safe use.
Batteries are not user serviceable.
Equipment protected from AC power by double insulation or reinforced insulation
by use of the supplied power supply. Will not cause electromagnetic interference
and complies with 601-1-2.
.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Operating:
Storage
:
Permissible environmental conditions during normal use.
Temperature
Humidity
Barometric Pressure
+18°C to +30°C
30-85% RH
100kpa-50kpa
-20°C to +45°C 10-85% RH 100kpa-50kpa