![dentalfarm MICRA Evoluzione Скачать руководство пользователя страница 3](http://html.mh-extra.com/html/dentalfarm/micra-evoluzione/micra-evoluzione_user-and-maintenance-manual_2483389003.webp)
3. INSTALLATION INSTRUCTIONS
/!\
ATTENTION
Installation of this machine is quite easy but must be carried out paying
utmost attention in order to avoid any mistake which may originate
problems, inconveniences and even damages during operation.
1. Place the machine on a proper workbench, which must be stable and strong enough to
hold the machine safely. Keep a sufficient distance (10/15 cm) on the right side of the
machine for the pipes and the feeding cables, as well as in order to have access to the
shutter and the filter. If a built-in WAFIS system is present, preferably choose a
workbench and a position (for instance on the bench edge) allowing the exhaust pipe to
fall perpendicularly into the collection tank.
2. Insert the quick clutch fitting on the male intake (A) located on the right side of the unit
and connect the pneumatic feeding pipe (polyethylene or rilsan, with diameter
∅
8x6)
to the pipe-fitting. It is also possible to use a
∅
12x6 elastic feeding pipe, by using the
fitting supplied with the unit.
3. Connect the electric feeding cable to the pre-fitted electric socket on the unit (B) and
plug into an approved 220v AC - 50Hz network socket with ground connection.
4. Install the preferred suction system; WAFIS, PRO-3 or any other.
The built-in WAFIS should be installed on the top left side of the
unit: unscrew the 4 screws fixing the black collection intake and
mount the WAFIS, unscrew the cap (C) and replace it by the pipe-
fitting delivered in the package, put together with the other portion
of pipe and plug the feeding cable into the electric socket (D).
PRO-3 suction units can be connected in two different ways:
on
the top left
to extract dust
particles only; in this case,
insert the piping into the black
collection intake; alternatively,
on the rear bottom
of
the working chamber:
in this case, all the
exhausted abrasive will
be extracted. In both
cases, remove the protection caps from their holes.
Any other suction system can be connected to the black
collection intake by means of flexible piping; the inside
manifold fits a
∅
30mm pipe, the outside manifold fits a
∅
40mm pipe.