Allways unplug the device before opening.

Alterations are only allowed after consultation and release of the manufacturer. 
In case of nonobservance any guarantee and any liability expires.

Any intervention of non authorised persons will cancel the validy of your 
guarantee

Working instructions :

1.To unscrew the safety stopper of the safety valve 

– press down and turn anti - clockwise

2.Fill up the tank with about 1,3 l of clean water, use only clean containers and clean funnel.
3.To fasten the safety stopper and the safety valve, press down and turn clockwise
4.Connect the cable to the mains ( 220 / 240 V ) and make sure that the voltage is the same as 
mentioned beneath your Denta Dampf 1500 AS
5.Press both switches – both lights will go on. This indicates that the machine is heating. If 
the right temperature is reached and the pressure of approximate 4 bars has been reached.
6.Now the machine is ready for use.
7.Pressing the switch on the handpiece – the steam will come out. Never steam to any 
electrical machinery or to people.

Never unscrew the safety stopper and safety valve while 
Denta Dampf 1500 AS is under pressure.
Never fill up with detergent products or others than clean 
water.

To fill up tank again during a cleaning operation proceed as follows :

1. Turn of the heating switch but not the main switch
2. Press the handle switch for complete evacuation of the steam inside the boiler
3. Disconnect the cable
4. Unscrew the safety stopper and the safety valve
5. Wait 2 or 3 minutes and fill up slowly 

– use hot water

To ensure the longevity of your Denta Dampf 1500 AS we recommend to descale the boiler at 
least once or twice a quarter. Proceed as follows :

1. Fill up 1 l of the tank with water and add a spoon full of descaler ( you`ll find in all 

supermarkets)

2. Close the tank and press the switches.
3. Wait about 15 minutes and disconnect the cable.
4. Let the descaler work overnight
5.  The next day unscrew the safety stopper and the safety valve , turn the steamer around            

and let the water flow out through the top. Then rince twice with cold water.

- 4 -

Содержание Denta Dampf 1500 AS

Страница 1: ...afety directions and observe them carefully before installing and commissioning the device After having unpacked the device please check carefully weather transportation damages of any kind have occurred If that is the case the carrier or manufacturer must be notified within 3 days The company will not be responsible for any damage found after that time Dental Kiefer Hüttenwerkstr 24 D 66763 Dilli...

Страница 2: ...built in aluminium boilers base Manostat tarred 4 bars 58 psi Electrovalve 12 V Silicon sheated cables 2 On off switches Safety stopper Hand piece with 1 4 m safety tube Height width depth 29 cm 33 cm 33 cm Tank volume gross 1 5 litres Noise level 60 db A Security Pressure controller Thermostatic control Safety valve Electrovalve 12 V 10 bars 144 psi Silicon sheated cables Length of cables accordi...

Страница 3: ...is damaged or has no more water in the tank To avoid unnecessary risks never dismantle the machine Call your dealer to get it verified or repaired It may not be filled with pressure and may not be connected to the water supply When you are working with other persons at your side increase your vigilance Avoid to connect several machines with power at the same electric circuit For your safety you mu...

Страница 4: ...switch on the handpiece the steam will come out Never steam to any electrical machinery or to people Never unscrew the safety stopper and safety valve while Denta Dampf 1500 AS is under pressure Never fill up with detergent products or others than clean water To fill up tank again during a cleaning operation proceed as follows 1 Turn of the heating switch but not the main switch 2 Press the handle...

Страница 5: ... pressure controller No steam Water in the Tank Dirt in the electrovalve decalcify device Electrovalve or transformator or handpiece damaged ask our after sales service Single parts list D 1403 pressure gauge complete 33 45 D 0756 switch on off 4 20 M 0093 safety cap complete 26 00 D 0824 electrovalve 12 V 49 80 D 0385 pressure buton 48 00 D 1052 transformator 220 230 12 V 26 00 M 0051 handpiece c...

Страница 6: ...r are not covered by this guarantee We expressly refuse any further liability Furthermore we refuse liability caused by improper use This guarantee becomes void when the purchaser or a third person carries out any alternations on the device Dental Kiefer Hüttenwerkstr 14 D 66763 Dillingen Service address Dental Kiefer Lortzingstr 1 a D 76694 Forst After your purchase please fill out and send the f...