5.   Securing Declination Angle 

Securing the declination angle will allow you to flip-up your loupe optics so that they 
pivot only on the upper joint.

5.1   To do so, securely tighten the lower hinge joint with a Phillips screwdriver. Increase or 

decrease the tension by tightening or loosening the upper screw.

5.2  Declination angle will be secured when optics are flipped up

5.1

5.2

5.1

5.2

6.  Cleaning and Disinfecting Lenses and Shields 

 

 Use the supplied cloth to clean the optical lenses and shields. A lightly moistened 
cloth with alcohol or eyeglass lens cleaner can also be used. DO NOT WET OR 
IMMERSE THE OPTICAL LENSES IN WATER OR LIQUID. For all other components 
such as the frames, nose pad and hinges, clean with an alcohol or disinfectant wipe. 
It is recommended that lens covers are used on the front of the loupes. These can be 
removed, cleaned thoroughly, dried and replaced on the loupes.

•  DO NOT SUBMERGE PRODUCT OR SUBJECT TO SURFACE SATURATION WITH ANY  

LIQUID OR ULTRASONIC CLEANER. Your product is not waterproof.

•  DO NOT CLEAN VIA AUTOCLAVE, CHEMICLAVE, GLUTARALDEHYDE, IODOPHOR, OR 

ANY OTHER STERILIZATION METHOD. 

• NO MODIFICATION OF THIS EQUIPMENT IS ALLOWED

7. 

 Storage 

When not using your loupes, store in the supplied carrying case with the lens covers 
in place (Legacy optics only). Avoid moisture and store in a dry environment.

denmat.com

Содержание PeriOptix Flip-up

Страница 1: ... between the hours of 8am to 5pm Pacific Time Monday Friday Outside the U S please call 1 805 347 7990 Microline Extremely lightweight and exceptional clarity Legacy Greatest value flip up on the market Panoramic Highest rated flip ups with a wider field of view Den Mat Holdings LLC 1017 W Central Avenue Lompoc CA 93436 USA denmat com MDSS GmbH Schiffgraben 41 30175 Hannover Germany ...

Страница 2: ...Flip up Loupe and Frame Directions for Use Instructions warranty and cleaning for PeriOptix Flip up loupe system and frame ...

Страница 3: ...Inspect contents If any item is missing contact DenMat Customer Support 1 Loupe Optics 2 Frame 3 Case 4 Lens Covers Legacy Optics only 5 Microfiber Cleaning Cloth 6 Anti Fog Cleaner 7 Glasses Strap 8 Phillips Screwdriver Legacy or Hex Screwdriver Panoramic 4 5 5 6 7 8 3 1 1 2 2 denmat com ...

Страница 4: ...glect accidents modification or shipping If the instrument is not used in accordance with manufacturer s recommendations or instructions If repaired or serviced by other than DenMat or a DenMat authorized representative Prompt inspection and reporting of missing or damaged product should be reported to shipping carrier and DenMat within 3 days of receipt of product If DenMat products or any compon...

Страница 5: ... be used DO NOT WET OR IMMERSE THE OPTICAL LENSES IN WATER OR LIQUID For all other components such as the frames nose pad and hinges clean with an alcohol or disinfectant wipe It is recommended that lens covers are used on the front of the loupes These can be removed cleaned thoroughly dried and replaced on the loupes DO NOT SUBMERGE PRODUCT OR SUBJECT TO SURFACE SATURATION WITH ANY LIQUID OR ULTR...

Страница 6: ...Loupe Terminology A Declination Angle B Convergence Angle C Interpupillary Distance IPD distance between the pupils A B C A C B denmat com ...

Страница 7: ...determine the working distance of the loupe optics is to view your thumbs while moving them in and out of focus 2 Adjusting IPD Interpupillary Distance 2 1 With the Optilock feature the only adjustment that is usually needed is to set the IPD Interpupillary Distance While focusing on an object pull up on the Interpupillary Adjustment Knob and rotate moving the optics apart and together Look at the...

Страница 8: ... step 2 Then follow the instructions in step 3 There is a convergence error if two horizontal dots are seen Rotate the lens barrels to eliminate the horizontal convergence error 4 1 To adjust the horizontal convergence angle first look downward at the loupe optics from the top to see if the barrels are equally angled 4 2 Barrels out of alignment will appear asymmetrical 4 3 For Legacy optics if a b...