0.1%

99.9%: ± ( 2.0% of rdg + 2 digits ), Frequency lower than 10kHz 

-- Sensitivity: sine wave 0.6V rms 
-- Overload protection: 500V DC or AC rms 

6-9 Temperature

 

Range 

Accuracy 

Resolution 

 

-20~150

 

± ( 3

+ 1digit ) 

1

 

150~1000

 

± ( 3% of rdg + 2digits ) 

 

-4~302

 

± ( 5

+ 2digits ) 

1

 

302~1832

 

± ( 3% of rdg + 3digits ) 

-- NiCr-NiSi K-type sensor 
-- Overload protection: 800mA/500V Fast Fuse

 

6-10 Diode and Audible continuity test 

Range 

Description 

Test Condition 

Display read approximately 
forward voltage of diode 

Forward DC current 
approx. 1.5mA 
Reversed DC voltage 
approx. 3.2V 

Built-in buzzer sounds if 
resistance is less than 50Ω 

Open circuit voltage 
approx. 1V 

Overload protection: 500V DC or AC rms

 

6-11 Transistor hFE test 

Test range: 0-1000 

Ib=10µA, Vce=1.8V Approx. 

6-12 Non Contact AC Voltage (NCV) detection 

Test voltage range: 90V

1000V AC rms 

The NCV red light and green light will light up alternately together with sound. 

7. OPERATING INSTRUCTIONS

7-1 Attention before operation 

7-1-1 Check battery. When the battery voltage drop below proper operation range, the “

” symbol 

will appear on the LCD display and the battery need to changed. 
7-1-2 Pay attention to the “

” besides the input jack which shows that the input voltage or current 

should be within the specified value. 
7-1-3 The range switch should be positioned to desired range for measurement before operation. 

7-2 Measuring DC & AC Voltage 

7-2-1  Set  the  rotary  switch  at  the  desired  “

mV

”  or  “

V

”  range  position,  it  shows  symbol  for 

testing DC voltage, if you want to test AC voltage, push “SELECT” button switch. 
7-2-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-2-3 Connect test leads across the source or load under measurement. 
7-2-4 You can get reading from LCD. The polarity of the red lead connection will be indicated along 
with the DC voltage value. 

NOTE:

   

1.“

”  means  you can’t  input  the  voltage  more  than  1000V  DC  or  750V  AC,  it’s  possible  to  show 

higher voltage, but it may destroy the inner circuit or pose a shock. 
2. Be cautious against shock when measuring high Voltage.

7-3 Measuring DC & AC Current 

7-3-1  Set  the  rotary  switch  at  the  desired  “

uA

”  &  “

mA

”  &  “

A

”  range  position,  it  shows 

symbol for testing DC current, if you want to test AC current, push “SELECT” button switch. 
7-3-2 Connect the black test lead to 

COMT-

 jack and the red to the 

μAmAT+

 jack for a maximum 

600mA current , for a maximum 6A or 20A current, move the red lead to the 

20A 

jack. 

7-3-3 Connect test leads in series with the load under measurement. 
7-3-4 You can get reading from LCD. The polarity of the red lead connection will be indicated along 
with the DC current value. 

NOTE: 

1. When  the  value  scale  to  be  measured  is  unknown  beforehand,  set  the  range  selector  at  the 
highest position. 
2. When  only  “OL”  is  displayed,  it  indicates  over-range  situation  and  the  higher  range  has  to  be 
selected. 
3. “

” means the socket mA’s maximum current is 800mA and 20A’s maximum current is 20A, over 

800mA or 20A current can be protected by the fast fuse. 
4. On the 20A range, the measuring time should be less than 10 seconds to prevent precision from 
affecting by circuit heating. 

7-4 Measuring DC & AC 600A Current 

7-4-1 Set the rotary switch at the desired “

600A

” range position, it shows symbol for testing DC 

current, if you want to test AC current, push “SELECT” button switch. 
7-4-2 Connect the black banana plug of the AC/DC Current Clamp Adapter to 

COMT-

 jack and the red 

banana plug to the 

μAmAT+

 jack. 

7-4-3 Set the AC/DC Current Clamp Adapter to “

1mV/A

” range. 

7-4-4 When perform DC current measurement, always rotate or push the DCA zero adjuster on the 
Clamp Adapter until the multimeter reads zero. 
7-4-5 Clamp the Jaws around the 

one 

conductor to be measured. Center the conductor within the Jaw 

using the Centering Marks as guides. 
7-4-6 Read the result from the LCD panel. The arrow in the Jaw indicates the DC current direction of 
positive current flow (positive to negative). 

7-5 Measuring Resistance 

7-5-1 Set the rotary switch at the desired “

” range position. 

7-5-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-5-3 Connect test leads across the resistance under measurement. 
7-5-4 You can get reading from LCD. 

NOTE: 

Max. input overload: 500V rms

10sec

 

1. For measuring resistance above 1MΩ, the mete may take a few seconds to get stable reading.
2. When  the  input  is  not  connected,  i.e.  at  open  circuit,  the  figure  ‘OL’  will  be  displayed  for  the
over-range condition. 
3. When checking in-circuit resistance, be sure the circuit under test has all power removed and that 
all capacitors have been discharged fully. 

7-6 Measuring Capacitance 

7-6-1 Set the rotary switch at the desired “

” range position, push “SELECT” to choose 

capacitance

 

measurement. 

7-6-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-6-3 Connect test leads across the capacitance under measurement. 
7-6-4 You can get reading from LCD. 

NOTE: 

Max. input overload: 500V rms

10sec

 

1.

Capacitors should be discharged before being tested.

2.

When testing large capacitance, it will take longer time before the final indication ( For 
100uF~99.99mF range, it will take about 10 seconds ). 

3.

When testing small capacitance (≤1uF), to assure the measurement accuracy, first press "REL

", 

then go on measuring. 

7-7 Measuring Frequency & Duty cycle 

7-7-1 Set the rotary switch at the desired “

Hz

” range position. 

7-7-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-7-3 Push “

Hz/Duty

” key to choose 

Frequency 

or 

Duty cycle 

test. 

7-7-4 Connect the probe across the source or load under measurement. 
7-7-5 You can get reading from LCD. 

7-8 Measuring Temperature 

7-8-1 Set the rotary switch at the desired “

” range position, push “SELECT” to choose 

 

or

 

 

measurement. 
7-8-2 Connect the black banana plug of the sensor to 

COMT-

 jack and the red banana plug to the 

μAmAT+

 jack. 

7-8-3 Put the sensor probe into the temperature field under measurement. 
7-8-4 You can get reading from LCD. 

NOTE:

 

1. The accessory of the meter WRNM-010 type contact thermocouple limit temperature is 250 

 (300 

 

shortly), please use special probe for test higher temperature. 
2. Please don't change the thermocouple at will, otherwise we can't guarantee to measure accuracy.
3. Please don’t importing the voltage in the temperature function.

7-9 Diode & Audible continuity Testing 

7-9-1 Set the rotary switch at the “

” range position, push “SELECT” to choose 

Diode 

or 

Audible continuity 

measurement. 

7-9-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-9-3 On 

diode

 range, connect the test leads across the diode under measurement, display shows 

the approx. forward voltage of this diode. 
7-9-4 On 

Audible continuity

 range, connect the test leads to two point of circuit, if the resistance is 

lower than approx. 50Ω, the buzzer sounds. 

NOTE: 

Make  sure  the  power  is  cut  off  and  all  capacitors  need  to  be  discharged  under  this 

measurement. 

7-10 Transistor hFE Test 

7-10-1 Set the rotary switch at the desired “

hFE

” range position. 

7-10-2 Determine whether the transistor is NPN or PNP and locate the Emitter, Base and Collector 
leads. Insert the leads into the proper holes in the socket on the front panel. 
7-10-3 You can get reading of the approximate hFE value from LCD. 

NOTE: 

Don’t connect an external voltage to measuring terminals. 

7-11 Non Contact AC Voltage detection 

7-11-1 Set the rotary switch at the desired “

NCV

” range position, the NCV green LED light will light up. 

7-11-2 Hold the Meter so that the mater’s top is vertically and horizontally centered and contacting the 
conductor, when the live voltage ≥ 90V AC rms, the NCV red LED light and green LED light will light 
up alternately together with sound. 

NOTE: 

1. Even without LED indication, the voltage may still exist. Do not rely on non-contact voltage detector 
to  determine  the  presence  of  voltage  wire.  Detection  operation  may  be  subject  to  socket  design, 
insulation thickness and different type and other factors. 
2. When the meter input terminals presence voltage, due to the influence of presence voltage, voltage 
sensing indicator may also be bright. 
3. Keep the meter away from electrical noise sources during the tests, i.e., florescent lights, dimmable 
lights,  motors,  etc..  These  sources  can  trigger  Non-Contact  AC  Voltage  detection  function  and 
invalidate the test. 

8. Battery replacement

8-1 When the battery voltage drop below proper operation range the "

 

" symbol will appear on the 

LCD display and the battery need to changed. 
8-2 Before changing the battery, set the selector switch to “

OFF

” position and remove the test leads 

from the terminals. Open the cover of the battery cabinet by a screwdriver. 
8-3 Replace the old battery with the same type battery (

6F22

 

9

V). 

8-4 Close the cover of the battery cabinet and fasten the screw. 

9. Fuse replacement

9-1 This meter is provided with a 800mA/500V fast fuse to protect the temperature test and the current 
measuring circuits which measure up to 600mA, with a 20A/500V fuse to protect the 20A range. 
9-2 Ensure the meter is not connected to any external circuit, set the selector switch to “

OFF

” position 

and remove the test leads from the terminals. Open the cover of the battery cabinet by a screwdriver. 
9-3 Replace the old fuse with the same type and rating: 6×30mm 800mA/500V fast fuse or 6×30mm 
20A/500V fast fuse. 
9-4 Close the cover of the battery cabinet and fasten the screw. 

10. Maintenance

10-1 You must replace the test leads if the lead is exposed, and should adopt the leads with the same 
specifications as origin. 
10-2 Use only moist fabric or small amount of detergent but not chemical solution for cleaning. 
10-3 Do not use the meter before the back cover is properly closed and screw secured.    Upon any 
abnormality, stop operation immediately and send the meter for maintenance. 
10-4 Please take out the battery when not using for a long time. 

11. Accessories

[1] Test Leads: electric rating 1000V 20A 
[2] “K” type thermocouple sensor probe

 

[3] Operator’s Manual 