4

4.  Changing the Mode 

 

Press the 

SET

 button twice. F2 will display on the screen. 

Press the left or right arrow to select the preferred 

MODE 

(BODY, SURFACE

 or 

COUNT)

. To save the setting, press 

the 

SET

 button until the 

HOME SCREEN

 is displayed. The 

new setting will appear on the LCD. (Figure 10)

5.  Setting the Alarm Temperature
 

Press the 

SET

 button three (3) times. F3 will display on the 

screen. Press the left and right buttons to adjust the alarm 
temperature between 98.6°F – 100.4°F (37.0°C – 38.0°C). 
The default alarm setting is 100.4°F (42°C). To save the 
setting, press the 

SET

 button until the 

HOME SCREEN 

is 

displayed. 

 

NOTE:

 

The alarm only works in the BODY mode; the alarm cannot be set in the 

SURFACE or COUNT modes. (Figure 11)

6.  Setting the Volume:

 

 

Press the

 SET

 button four (4) times. F4 will display on the 

screen. Press the right arrow to increase the volume. Press 
the left arrow to decrease the volume. 

As the volume is 

increased, the speaker symbol will expand, and the 
beep tone will become louder. (Figure 12) To mute the 
volume, press the left arrow until the speaker symbol 
no longer appears. (Figure 13) To save the setting, press 
the SET button until the HOME SCREEN is displayed. 

7.  Setting the Backlight
 

Press the 

SET

 button five (5) times. F5 will display on the 

screen. Press the left arrow or right arrow to turn the backlight 

ON 

or 

OFF

. To save the setting, press the 

SET 

button until 

the 

HOME SCREEN

 is displayed. (Figure 14)

8.  Setting the Language: 

 

The device utilizes voice commands.

 To select the 

preferred language, press the 

SET

 button six (6) times. F6 will 

display on the screen. Press the left arrow or right arrow to 
select either 

CH

 for 

Chinese

 or 

EN 

for 

English

. To save the 

setting, press the 

SET

 button until the 

HOME SCREEN 

is 

displayed. (Figure 15)

9.  Setting the Output Signal

 

Press the 

SET 

button seven (7) times. F7 will display on the 

screen. Press the left arrow or right arrow to select the output 
type. Select either S1 for 

SWITCH

 mode or S2 for 

PULSE

 

mode. To save the setting, press the 

SET

 button until the 

HOME SCREEN

 is displayed. (Figure 16)

Body Mode

F2

Fig. 10

Fig. 11

UOL

0 1

Fig. 13

UOL

0 1

Fig. 12

UOL

0 1

Fig. 14

UOL

0 1

O

f5

U

C

f6

H

Fig. 15

5

f7

1

Fig. 16

Содержание ThermoTrace 15053

Страница 1: ...User Guide Auto Check Non Contact Infrared Forehead Thermometer K3 Pro Model 15053 Please read this manual before operating unit Important safety information inside...

Страница 2: ...ps www deltatrak com 15053 2 Support and Technical Services https www deltatrak com support tech support 3 Warranty https www deltatrak com support warranty 5 Introduction Video https www deltatrak co...

Страница 3: ...ting Hole 11 Battery Cover Accuracy 0 2 degrees 93 F 113 F 34 C 45 C Alarm Mode Flashing Red LED audible tone for 10 seconds Measuring Distance 2in 6in 5cm 15cm Display LCD Transmission Interface USB...

Страница 4: ...Saved Temperature Reading or Environmental Temp Ambient Temperature 7 Environmental Temperature Reading Normal Temperature When the temperature reading is normal green lights flash on both sides and t...

Страница 5: ...y History Temp and Environmental Temp The saved temperature log number i e M01 will also be displayed Press the left arrow to scroll through the saved readings in the order that the measurements were...

Страница 6: ...ome louder Figure 12 To mute the volume press the left arrow until the speaker symbol no longer appears Figure 13 To save the setting press the SET button until the HOME SCREEN is displayed 7 Setting...

Страница 7: ...nnection Use Type C USB cable to connect to your computer Download the software and connect the device in order to store measurement results in real time For more online instructions please refer to t...

Страница 8: ...The subfolder TempData year month is automatically generated within the TestACY folder The format is Year Month Day The file type is TXT If you cannot find the folder use your computer s search functi...

Страница 9: ...Temp 99 5 Bod Save As 97 5F Normal COM V1 03 Serial Port Operation Disconnect COM1 Quantity 1 15 12 30 97 5F 74 3F 1 15 12 32 97 7F 74 3F Test Time Test Value Environment TESTACY E success Jul 6 2020...

Страница 10: ...e data click Save as Name the file and select the file type txt or csv Set the upper limit of the alarm When the temperature reading exceeds this value the thermometer will give an audible alarm Quant...

Страница 11: ...The ThermoTrace Auto Check Pro Non Contact Infrared Forehead Thermometer was designed for long term professional use and has a limited warranty period of 1 year from the date of purchase against defec...

Страница 12: ...64 8923 2778 www deltatrakchina com cn DTI China Limited NT Hong Kong 852 3568 5538 www dtiap com DTI Asia Pacific Auckland New Zealand 64 9 5757 886 www deltatraksouthpacific com DTI South Pacific Os...

Отзывы: