QUICK GUIDE LTLXL-UK VERS 1.01.DOC
2008-10-31
Calibration procedure
Before a calibration is carried out make sure that the calibration unit is clean.
Press the
CALIBRATION
button
. Select RL and/or Qd calibration. Mount the calibration unit underneath
the instrument. Make sure that the pins on the interface plate fit into the holes in the calibration unit.
Press the
OK
button to start calibration. Check the value displayed and, if necessary, adjust the calibration
value shown in the display so it matches the value stamped on the calibration unit.
Press the
OK
button to finish the calibration or repeat if erroneous.
The calibration procedure is now done. Remove the calibration unit and store it properly. Press
OK
to return to the
measurement display.
Miscellaneous
The measure display can at all times be activated by pressing
Reset Log:
Press the
MENU
button
and select
LOG/CLEAR DATA/ALL
.
Date and Time:
Press the
MENU
button and select
SETTINGS/DATE & TIME/SET TIME/DATE
. Select wanted
value and use
▲
and
▼
to set the time and date. Accept the setting by pressing the green
OK
button. Alternatively
SYNC TO GPS
can be selected to auto set the time. Remember
also to set correct time zone.
Power Save:
Press the
MENU
button and select
SETTINGS/POWER
. Select
AUTO SLEEP
or
AUTO OFF
and use
▲
or
▼
to edit the auto turn off / sleep time. Accept the setting by pressing the
green
OK
button.
Sound:
Press the
MENU
button
and select
SETTINGS/SOUND
.
SELECT KEY CLICK
or
SOUNDS
to set the individual sound levels for key click and miscellaneous system sounds.
Use
▲
or
▼
buttons to set the levels. Accept the setting by pressing the
OK
button.
Warnings and Errors: An error icon
or warning icon
will be displayed in the top line of the display
if a problem is detected when a measurement is taken. Press
▲
to get a description of the
most severe error or warning. Now press the
OK
button to display a total list of errors or
warnings.