SWT4 User Manual v3.1
Service and maintenance
33
4.2 Testing
4.2.1 Calibration
When delivered Tensiometers are calibrated with an offset of 0 hPa
(when lying horizontally) and a linear response. The offset of the
pressure transducer has a minimal drift over the years. Therefore, we
recommend you check sensors once a year and re-calibrate them
every two years.
Return the Tensiometers to Delta-T for recalibration, or contact us for
details about available calibration accessories.
4.2.2 Check the Offset
There are two ways to check the offset.
1. Connect the Tensiometer to a readout device. Fill the beaker with
distilled or de-ionized water and immerse the ceramic cup to a depth of
7.5 cm. Wait until the reading is stable. If there are bubbles inside the
cup this might take a while. The reading now should be
0 kPa.
2. To check the zero-point more precisely unscrew the cup. Shake the
pressure sensor to remove water from the pressure transducer hole.
The offset is acceptable when the reading is between -0.5 and +0.5
kPa.
The pressure sensor diaphragm is inside the small hole on the
pressure sensor body. It is very sensitive and must never be
touched! It can be destroyed even by slightest contact! No
contamination should get on the sealing and gasket.
Before reassembling cup and sensor body carry out the degassing
procedure (see chapter 4.1.2 "Refilling in the lab" on page 27).
For testing the gain a calibration kit is required. Contact Delta-T for
details