Installation 

Step 1: 

You can install one or two GP1-PBA1s in the GP1. Align the GP1-PBA1 with the 

CH1

 

or 

CH2

 terminals so that one of its 5 pins overhangs the GP1 terminal block. Insert the 

remaining 

four

 pins into the 

four 

screw terminals of 

CH1

 or 

CH2. 

Tighten the screws. 

Pin not

connected

CH2 (PWR)

 

Pin not

connected

CH1 (PWR)

 

SWT tensiometers 

Sensor connection 

Connect tensiometer wires to 
GP1-PBA1 terminals: 

Brown Vref 
White CH(-) 
Black CH(+) 
Blue GND 

Connect the screen (thick black) 
to any GND terminal on the GP1 
board. 

This connection method gives 
negative hPa readings for 
negative potentials (i.e. suction).  

 

SWT tensiometer connection to CH2 

Step 2: Measure the zero offset 

Before programming the GP1 for tensiometer readings you need to measure the 

zero offset

 of the GP1-PBA1. 

SWT3 

and 

SWT4 tensiometers: 

remove the transducer or carefully unscrew the 

ceramic cup from the tensiometer. Remove any water droplets from the access hole to 
the pressure transducer by gently tapping the unit. Connect the tensiometer to the 
GP1-PBA1.  

Refer to the tensiometer User Manual and carefully reassemble and refill the 
tensiometer after measuring the zero offset. 

 

 

©