Analog outputs (only PM[B]sense-A):
Command
Reply
Description
CA1On
&|
Enable/disable the offset of the analog output 1:
Offset disabled if n=0 (0…20 mA or 0…10 V)
Offset enabled if n=1 (4…20 mA or 2…10 V)
Default
: Offset enabled (n=1) if the output is current,
offset disabled (n=0) if the output is voltage (optional)
RA1O
& n|
Reads the setting of the offset for the analog output 1.
CA1SOn
&|
Sets the direct or inverse correspondence between analog
output 1 and associated physical quantity:
4 mA/0 V
Min. quantity, 20 mA/10 V
Max.
quantity if n=0
20 mA/10 V
Min. quantity, 4 mA/0 V
Max.
quantity if n=1
Default
: Direct correspondence (n=0)
RA1SO
& n|
Reads the type of correspondence (direct or inverse) be-
tween analog output 1 and associated physical quantity.
CA1Tn
&|
Associates the analog output 1 to:
PM1.0 if n=0
PM2.5 if n=1
PM10 if n=2
CO
2
if n=12 (only PMBsense-A)
Default
: PM2.5 (n=1)
The PM measurement is averaged according to what is set
with CPS command or MODBUS holding register 19.
RA1T
& n|
Reads the physical quantity associated to analog output 1.
CA1Ln
&|
Sets n as the minimum value of the measuring range of
the physical quantity associated to the analog output 1.
If the output is associated to PM, the value must be ex-
pressed as a number of tenths (e.g. n=5 to indicate 0.5
µg/m
3
). If the output is associated to CO
2
, the value must
be expressed as a number of units (e.g. n=200 to indicate
200 ppm).
Default
: 0
RA1L
& n|
Reads the minimum value of the measuring range of the
physical quantity associated to the analog output 1.
CA1Hn
&|
Sets n as the maximum value of the measuring range of
the physical quantity associated to the analog output 1.
If the output is associated to PM, the value must be ex-
pressed as a number of tenths (e.g. n=50 to indicate 5.0
µg/m
3
). If the output is associated to CO
2
, the value must
be expressed as a number of units (e.g. n=800 to indicate
800 ppm).
Default
: 10000 (=1000.0 µg/m
3
of PM)
RA1H
& n|
Reads the maximum value of the measuring range of the
physical quantity associated to the analog output 1.
RA1F
&
Quantity
Minimum value
Maximum value
Simultaneously provides the information obtainable with
the RA1T, RA1L e RA1H commands.