7. COMMUNICATION PROTOCOL
All models are supplied with a serial output with mini-USB connector, that allows you
to upload commands and to download data through your PC using communication
programs, such as – for example - Hyperterminal.
Connect the serial output located on the instrument side to a USB port of your PC via
the special RS45 or RS45I connection cable, which is supplied with a built-in USB
adaptor and mini-USB connector to the instrument end, and a USB connector to the
PC end. For a proper connection operation, it is necessary to install the drivers
included in the CD-ROM supplied with the instrument in your PC.
Attention: serial connection doesn’t work if the instrument is directly connected to the
PC via a standard mini-USB/USB adaptor cable.
After the connection, the instrument appears connected to a COM-type port.
Setup serial transmission parameters as follows:
•
Baud rate: 115200
•
Parity: None
•
Data bits: 8
•
Stop bits: 2
Please find below the list of configuration commands and data requests that can be
sent through a PC, with the instrument connected via mini-USB output:
Tab. 7.A: Communication protocol
Command
Response
Description
G0 HD…
Transmitter
model
G1
Vnn.nn aaaa/mm/gg
Firmware version and date
G2
aaaa/mm/gg hh.mm.ss
Calibration date and time
G3
nnnnnnnn
Instrument serial number
G4 nnnnnnnn Relative humidity module serial
number
G5 aaaa/mm/gg
hh.mm.ss
Relative humidity module factory
calibration date and time
GS
Download last logging session
GT
Download all logging sessions
PW
nnnnnn
Print number to retrieve access code
PWC nnnnnn nnnnnn &
Set new “nnnnnn” access code
PWnnnnnn USER
ENABLED! Upload access code and set
authorization level “1”
PWX
LOCKED!
Set authorization level “0”
RLnnn n
Print authorization level of parameter
number “nnn”
RPnnn
Print value of parameter number
“nnn”
S0
&
Disable measurement print
S1
&
Single measurement print
S2
&
Enable measurement continuous print
WLnnn x
&
Set authorization level of parameter
number “nnn” to “x” value
WPnnn val
&
Set parameter number “nnn” to “val”
value
For the available parameter list, refer to Tab. 5.A (page 29).