5)

  Press 

DATA/ENTER

 to proceed with the calibration.  The electrode offset mV value (OFFS) is 

shown on the display for few seconds.  
After a few seconds the instrument shows the measurement again, corrected according to the 
new calibration, 

but remains in calibration mode

Pressed repeatedly the 

DATA/ENTER

 key allows the calibration on the point to be repeated, for 

example, in order to obtain a more stable value. 

6)

 To end the electrode calibration press 

pH/mV (ESC

), or continue the calibration for the second 

point.  

7)

 Extract the electrode from the buffer, clean it carefully, and insert it in the following buffer. 

8)

 Press one time the 

°C/°F key (MENU)

9)

 The instrument displays the value detected on the new buffer: continue by repeating the steps 

from point 3. 

 

NOTES: 

 

After calibration, the instrument displays 

an electrode quality indication

-

 

no signal: electrode functioning. 

-

 

a small square blinking on the lower right: electrode almost exhausted. 

-

 

two small squares blinking on the lower right: electrode exhausted to be replaced.  

 

The 3 point calibration must always be carried out according to the fixed sequence: 
NEUTRAL>>ACID>>BASIC

. The basic buffer must be the last in the sequence. 

 

Without having pressed DATA/ENTER at all, the calibration is interrupted by pressing 
pH/mV (ESC);

 the previous values will continue to be used. 

 

The buffers are always presented in the sequence set in the menu with the parameters: 
-

 

BUFR_1, 

-

 

BUFR_2, 

-

 

BUFR_3.   

 

The 2 point calibration is possible using the sequence BUFR_1 - BUFR_2  or  BUFR_2 - 
BUFR_3  or even  BUFR_1 - BUFR_3. 

 

The 3 point calibration is only possible using the exact sequence BUFR_1 - BUFR_2 - 
BUFR_3

.   

 

 

Allowed calibration sequences

 

BUFR_1 BUFR_2 BUFR_3 

BUFR_1 - BUFR_2 

BUFR_2 - BUFR_3 

BUFR_1 - BUFR_3 

Number of  

calibra

tion  

points

 

BUFR_1 - BUFR_2 - BUFR_3 

 

 

In any case where no operation is performed, the “calibration” mode automatically stops 
after 60 seconds. 

 

If the calibration is rejected by the instrument because it is considered to be excessively 
corrupted, the 

CAL ERR

 message will appear, followed by a long beep. The instrument remains 

in calibration mode and maintains the previous calibration values: at this point, if the calibration 
is interrupted using 

pH-mV

/

ESC

, the instrument signals the anomaly through the CAL message 

blinking. 

 
 

Содержание HD2305

Страница 1: ...ation written in this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manua...

Страница 2: ...2 pH meter Thermometer HD2305...

Страница 3: ...ode increases the displayed nominal buffer value 9 C F MENU key it allows selection of the unit of measurement when pressed together with the DATA key it allows to open the menu 10 CAL starts the cali...

Страница 4: ...probes 15 4 3 3 Direct connection of 4 wire Pt100 sensors 17 5 WARNINGS 18 6 INSTRUMENT SIGNALS AND FAULTS 19 7 INSTRUMENT STORAGE 20 8 LOW BATTERY WARNING AND BATTERY REPLACEMENT 21 8 1 WARNING ABOU...

Страница 5: ...or three points at 4 01pH 6 86pH and 9 18pH The probes are fitted with the SICRAM automatic detection module with the factory calibration settings already being memorized inside The units of measureme...

Страница 6: ...ument ready for normal measurement The following is displayed MIN cm min s 2 3 m m k AUTO OFF the AutoPowerOff function can be disabled by simultaneously pressing this key and the C F MENU key when tu...

Страница 7: ...memorizing the MAX MIN and AVG values To reset the previous values and start with a new measurement session press DATA until the FUNC_CLRD message appears use the and arrows to select YES confirm by p...

Страница 8: ...o exit the menu press the C F key again For further information see chapter 3 Press this key to start the calibration procedure see par 4 2 pH electrode calibration The REL mV key is used for the foll...

Страница 9: ...hm Pt1000 probes are automatically detected by the instrument upon turning on the Probe Type menu item is configured by the instrument and cannot be modified by the user The temperature probes direct...

Страница 10: ...ature The probes are detected during turn on and this cannot be performed when the instrument is already on therefore if a probe is connected and the instrument is on it is necessary to turn it off an...

Страница 11: ...6mV pH T 0 C 54 20mV pH When a temperature probe is present the instrument automatically applies the ATC function automatic temperature compensation To disable it the temperature probe must be disconn...

Страница 12: ...one for the neutral and one for the alkaline band 25 C BUFR_1 NEUTRAL 6 860 BUFR_2 ACID 4 010 BUFR_3 ALKALINE 9 180 If electrode calibration was not carried out on the instrument or the batteries have...

Страница 13: ...point calibration must always be carried out according to the fixed sequence NEUTRAL ACID BASIC The basic buffer must be the last in the sequence Without having pressed DATA ENTER at all the calibrati...

Страница 14: ...C pH C pH C pH 0 6 98 50 6 83 0 4 01 50 4 06 5 6 95 55 6 83 5 4 00 55 4 07 10 6 92 60 6 84 10 4 00 60 4 09 15 6 90 65 6 85 15 4 00 65 4 10 20 6 88 70 6 85 20 4 00 70 4 13 25 6 86 75 6 86 25 4 01 75 4...

Страница 15: ...be inserted to a depth of at least 60 mm the sensor is housed in the end part of the probe NOTE when measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cav...

Страница 16: ...Used Caution Before welding pass the probe cable through the fairlead and gasket 6 Weld the wires as shown in the table Sensor Board connection Jumper to close Pt100 4 wires 4 JP4 3 JP3 1 JP1 2 JP2 Pt...

Страница 17: ...ve to be soldered as indicated in the figure on the left In order to use this type of probe it is necessary to set up the menu item Probe Type as described at page 9 The P100 probe is recognized upon...

Страница 18: ...ce measurements contact probes must be held perpendicular against the surface Apply oil or heat conductive paste between the surface and the probe in order to improve contact and reduce reading time W...

Страница 19: ...by the instrument is disconnected At the same time an intermittent beep is issued FUNC CLRD max min and average values cleared NEW_PROB_DET new probe detected OFFS offset OVER Measurement overflow th...

Страница 20: ...n places where humidity is high the instrument may be exposed to direct sunlight the instrument may be exposed to a source of high temperature the instrument may be exposed to strong vibrations the in...

Страница 21: ...nstrument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 scre...

Страница 22: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Страница 23: ...teries Power absorbed with instrument off 20 A Connections Input module for the temperature probes 8 pole male DIN45326 connector pH mV input Female BNC Unit of Measurement pH mV C F Measurement of pH...

Страница 24: ...ion 50 C 200 C 0 25 C 50 C 200 C TP472I 0 Immersion 50 C 400 C 0 25 C 50 C 350 C 0 4 C 350 C 400 C TP473P 0 Penetration 50 C 400 C 0 25 C 50 C 350 C 0 4 C 350 C 400 C TP474C 0 Contact 50 C 400 C 0 3 C...

Страница 25: ...ned pH electrode 3 diaphragms for milk cream etc gel filled with screw connector S7 body in glass Ag AgCl sat KCl KP62 Combined pH electrode 1 diaphragm for pure water paints etc gel filled with screw...

Страница 26: ...m length 230 mm contact surface 5 mm Cable length 2 metres TP475A 0 Pt100 sensor air probe Stem 4 mm length 230 mm Cable length 2 metres TP472I 5 Pt100 sensor immersion probe Stem 6 mm length 500 mm C...

Страница 27: ...27...

Страница 28: ...il pr sum d fectueux Dieser Garantieschein muss der Spedition beigelegt werden wenn das Ger t an das Kundendienstzentrum gesandt wird WICHTIG Die Garantie ist nur g ltig wenn dieser Abschnitt bis ins...

Отзывы: