10
3.1
C
OMBINED PROBE
LP
471
P-A
LP 471 P-A
is a two sensors combined probe with SICRAM module for measuring
illuminance
(lux) with standard photopic spectral response and
irradiance
(μW/cm
2
) in
UVA
spectral range
(315...400 nm, with peak at 365 nm). Moreover probe provides the ratio of UVA irradiance and il-
luminance in µW/lumen (quantity of interest in the museums field).
Both sensors are equipped with diffuser for the correction according to the cosine law.
At switching on, the instrument displays alternatively the measures of the two sensors. Pressing any
key (excluded ON/OFF key), automatic commutation is turned off. To reactivate it, press HOLD
and REL keys together.
To select the measure to display, press UNIT key.
3.2
C
OMBINED PROBE
LP
471
A-UV
EFF
LP 471 A-UVeff
is a two sensors combined probe with SICRAM module for measuring
total ef-
fective irradiance
according to the UV weighting curve. Two sensors are used to correctly measure
the total effective irradiance in the range 250...400 nm.
Both sensors are equipped with diffuser for the correction according to the cosine law. The probe
provides the total effective irradiance (“Er”
at display), the effective irradiance in the range UV-CB
(“BC” at display) and the UVA irradiance (“A” at display).
At switching on, the instrument displays alternatively the measures of the two sensors. Pressing any
key (excluded ON/OFF key), automatic commutation is turned off. To reactivate it, press HOLD
and REL keys together.
To select the measure to display, press UNIT key.
3.3
P
ROBES
LP
471
PYRA
02
AND
LP
471
PYRA
03
LP 471 PYRA 02
and
LP 471 PYRA 03
probes measure the
global solar radiation
in the spectral
range 300…3000 nm. They are composed of a first class (LP PYRA 02) or a second class (LP
PYRA 03) pyranometer and a cable with SICRAM module.
Since calibration data of pyranometer are saved in SICRAM module, the cable cannot be used to-
gether with other pyranometers.
The
global solar radiation
is expressed in W/m
2
or in μW/cm
2
. Measuring range is 0…2000 W/m
2
.
The module is equipped with a 5 m or 10 m long cable ending with a 4 poles male connector to be
inserted in the female connector of the pyranometer.
No user calibration is required.
The SICRAM module is recognized at the instrument switching on, then it’s absolutely necessary to
connect the module before turning the instrument on
.
3.4
P
ROBE
LP
471
SILICON-PYRA
LP 471 Silicon-PYRA
measures the
global solar radiation
using a silicon photodiode in the spec-
tral range 400…1100 nm.
The diffuser allows a correction according to the cosine law.
The probe is suitable for the measurement of natural sunlight. Under overcast skies, or for reflected
light measurements the use of a thermopile pyranometer is recommended (LP 471 PYRA 03 or LP
471 PYRA 02).
Содержание HD2302.0
Страница 2: ... 2 Photo Radiometer HD2302 0 ...
Страница 31: ... 31 ...