POLAROGRAPHIC PROBE
1.
Remove the protective container from the head of the probe.
2.
Unscrew the probe's head with the membrane permeable to oxygen.
3.
If necessary, replace the membrane.
4.
Fill the head of the probe with the supplied electrolyte solution.
5.
Eliminate any possible air bubble in the electrolyte solution.
6.
Screw the probe's head back on carefully.
DO9709SG
GALVANIC PROBE
1.
Remove the protective container from the head of the probe.
2.
Unscrew the cap of the filling hole and the head of the probe with the membrane permeable to
oxygen (check that the membrane is in good condition).
3.
Fill the head of the probe up to ¾ with KOH electrolyte solution.
4.
Screw the head of the probe with the electrolyte solution to the probe. Rinse any excess
electrolyte solution with water.
5.
Pour the electrolyte solution in the filling hole (the probe filling requires about 5 ml of
solution). From time to time, gently tap the probe stem to remove any air bubbles.
6.
When the filling hole is full, screw the cap back on. Rinse any excess electrolyte solution with
water.
7.
Place the probe with the membrane upwards and check that there are no air bubbles under the
membrane. If you notice air bubbles, further fill the probe with electrolyte solution.
Head of the probe
Membrane
Head of the probe
Filling hole