background image

4-pole M12

male connector 

of the datalogger 

(visible side)

CP2013-DB cable

Connector

Function

Color

1

Rain gauge contact

Brown

2

Rain gauge contact

White

3

Not connected

Blue

4

Not connected

Black

Fig. 3: Connection of the CP2013-DB cable to the rain gauge

Over-voltage protection

: the data loggers supplied with the mast clamping already mounted 

on   the   back   of   the   housing   are   provided   with   internal   over-voltage   protection   devices, 
connected to the clamping. For the correct operation of the protection devices, the mast on 

which the clamping is fixed, must be connected to ground.

Keyboard

The two function keys that allow to control the instrument are formed by two reed contacts, 

operable from outside by means of a magnet supplied with the instrument.
Two red areas are indicated on the front of the instrument, corresponding to the word "ENTER" 

and   to   the   symbols  

  and  

:   on   these   areas  will   be   placed   the   magnet  to   operate   the 

instrument. Keeping the magnet on the red area for a few instants and then moving it away is 

equivalent to press a button. For simplicity, in the description of the instrument that follows, 
this operation will be called: "

press the ENTER button

", "

press the 

/

 button

".

After use, the magnet with its support should be placed back into its holder.

Operation

The data logger counts and stores the number of emptying of the rain gauge tipping bucket. 

Each emptying of the bucket corresponds to a quantity of rain equal to  the resolution of the 
rain   gauge.   The   resolution   can   be   set   from  

0.050   to   1.599   mm   of   rain   by   using   the 

HD32MTLogger   software

.   The   rain   gauge   contact   can   be   both   NC   and   NO   type   (it   is   not 

necessary to set the contact type, the data logger automatically  works with both types of 

contact).

Status indicator

: the LED indicator of the data logger indicates the status of the contact of 

the rain gauge. The LED blinks red if the contact is open, and green if the contact is closed. 

Therefore, the LED color allows to highlight the emptying of the bucket and check 

that the bucket returns in the stable position after emptying

.

Display

: the display is backlit; to minimize the battery consumption, the display is switched 

off if no operation is peformed on the data logger for 2 minutes. However, the data logger is 
still working even when the display is off, and its activity is signaled by the flashing status 

indicator LED. In order to turn the display on, press any of the two buttons. When the display 
is on, the following screens are displayed sequentially:

3

Отзывы: