ERROR INDICATION
OL: Overload warning.
E1: The solution used for conductivity calibration is wrong because the
slope calculation gives a value >150% of the rated value.
E2: The solution used for conductivity calibration is wrong because the
slope calculation gives a value <30% of the rated value.
E3: The temperature of the calibration solution is not between 15°C
(59°F) and 30°C (86°F). Calibration is inadmissible outside this range.
ENGLISH
– 45 –
(mS•cm
-1
)
10
2
10
5
10
3
10
6
10
4
10
7
10
10
4
0.1
1
10
100
0.1
1
10
100
Resistance (
Ω
•cm)
Conductivity (
µ
S•cm
-1
)
Correspondence between electric resistance and conductivity