Troubleshooting 

 

No. 

Symptom 

Cause 

Action 

Tester does not start 

No/very discharged 
batteries 

Check batteries 

Electrical malfunction 

Replace the tester 

Display shows “AM2 
not found” 

No AM2 unit is 
present 

Insert an AM2 in the bracket and press 
Test 

AM2 unit is not 
inserted properly 

Make sure the AM2 unit is inserted 
according to 

Figure 1

AM2 unit is in 

back-off 

mode

 

Let the AM2 unit rest for at least 5 
minutes at a safe distance (>5 m / 17 
feet) from any ID-equipment such as sort 
gates, ID-readers, feeding stations, etc 
and try again. 

Radio interference 

Wait 5 minutes and try again in an area 
with less radio traffic 

AM2 unit is out of 
battery or faulty 

Replace the AM2 unit 

Tester is faulty 

Try with a couple of other AM2 units. If 
not working, replace the tester 

Display shows “Low 
battery” 

Discharged batteries 

None needed immediately but batteries 
will be needing replacement shortly 

Low operating 
temperature 

Move to a warmer location 

Display shows 
“Replace battery!” 

Very discharged 
batteries 

Replace batteries 

 