10
Setting the date
Setting the date
Setting the time
Setting the time
Exit
Return to the main menu
Calibration
a)
Connect fresh air to the sensor input. The icon
„
“
flashes
.
After the value stabilizes, move onto the next step using
„ “
b)
Using „ “ select the concentration of the calibration gas, confirm
„
“
c)
Connect the calibration gas to the sensor input. The icon
„
“
flashes
. Wait until the value stabilizes and confirm
„
“
d)
Using „ “ switch between „YES“ - save calibration, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
e)
In case of saving a new calibration, the transmitter will restart.
Setting the alarms
a)
Using „ “ select the concentration for alarm 1. Confirm „ “
b)
Using „ “ select the concentration for alarm 2. Confirm „ “
c)
Using „ “ select the concentration for alarm 3. Confirm „ “
d)
Using „ “ select the concentration for alarm 4 Confirm „ “
e)
Using „ “ switch between „YES“ - save settings, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the range of the 20mA current loop
a)
Using „ “ select the concentration appropriate to 20mA range. Confirm „ “
b)
Using „ “ switch between „YES“ - save settings, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
Changing the password
a)
Using „ “ select a number in thousands. Save the selected number „ “
b)
Using „ “ select a number in hundreds. Save the selected number „ “
c)
Using „ “ select a number in tens. Save the selected number „ “
d)
Using „ “ select a number in units. Save the selected number „ “
e)
Using „ “ switch between „YES“ - save settings, or „NO“ - return to the Settings menu. Confirm the selected
option „ “
Setting the year
a)
Using „ “ select a number in thousands. Save the selected number „ “
b)
Using „ “ select a number in hundreds. Save the selected number „ “
c)
Using „ “ select a number in tens. Save the selected number „ “
d)
Using „ “ select a number in units. Save the selected number „ “