16
Add-on modules
Product code
Name
Product description
20200004
DEGA NB III Relay Modul
Internal 2-relay, 250 V/10 A
Attachments
Table of error codes
code
cause
solution
1
Sensor is not present (EEPROM is not
communicating)
Disconnect and reconnect the sensor,
then restart the sensor by disconnecting
and reconnecting the power
2
Unknown sensor type
Contact the manufacturer
3
Type 2 sensor ID not found in the settings
table LMP91000
Contact the manufacturer - FW update
required
4
the EEPROM sensor CRC does not match
Sensor error - contact the manufacturer
5
LMP91000 not responding
Sensor error - contact the manufacturer
6
The set sensor ID does not match the
connected sensor
Checking the configuration of the sensor
required
7
Reserved
Reserved
8
EL sensor test error
Sensor error
9
Detector is in preheating mode
Wait a few minutes, the detector will
automatically enter measurement mode
10
Reserved
11
reserved
12
Error reading the internal FLASH
Restart the detector. If the error persists,
contact the manufacturer
13
Error reading the internal FLASH
Restart the detector. If the error persists,
contact the manufacturer
14
Error reading the internal FLASH
Restart the detector. If the error persists,
contact the manufacturer
15
Sensor error
Replace the sensor
16
The device exceeded the maximum
calibration interval by 50 %
Calibration is necessary
129
The measured value exceeds the range of -
10 %
If the error persists, contact the
manufacturer
130
The result of the AD conversion is out of
range
Contact the manufacturer