Do not open the transmitter and do not replace the sensors in the transmitter in a potentially explosive atmosphere 

 

Opening the cover and replacing the sensor in the transmitter in a potentially explosive atmoshpere can cause an explosion. If 
service  is  necessary,  first  unplug  the  device  from  the  power  supply.  After  that  the  device  may  be  disassembled  and  the 
sensors may be replaced (certified partner or manufacturer). 

Do not disassemble the product and ensure against it's contact with water 

 

Contact with internal components of the product may cause an electric shock. In case of any malfuction entrust the servicing 
of  the  product  exclusively  to  a  certified  service  centre.  Contact  with  water  can  create  a  short  circuit  in  the  product  and 
consequent damange to property or personal injury. 

Use appropriate cable types 

 

To ensure compliance with the parameters of the product, only use cable types recommended in this guide. 

Dispose of used products and trasmitter sensors with respect to the environment 

 

Transmitter  sensors  contain  hazardous  substances.  Dispose  of  them  in  accordance  with  the  current  legislation  on 
environmental protection. 

Use the transmitter only with the appropriate certified DEGA products 

 

The  device  is  certified  as  functionally  and  technically  qualified  only  with  original  "DEGA"  accessories.  In  case  of  using  the 
device with any other products the manufacturer is not liable for any damages that may occur.  

Undertake regular functional checks and calibrations of the transmitter 

 

Perform  regular  "CALIBRATION"  (setting  the  detection  limits,  checking  the  responsivness  of  the  sensor,  checking  the 
functionality  of  the  transmitter)  and  "OPERATIONAL  AND  FUNCTIONAL  CHECKS"  of  the  entire  detenction  system  (sensor 
excitation  with  subsequent  control  of  optical  and  audible  alarms,  triggering  fans,  shutdown  technology,  etc.).  Perform 
calibration and operational and functional checks only at certified service centers with a valid certificate of competence or the 
manufacturer. 
 

Warning: The transmitter automatically checks it's calibration period - the period of validity of it's calibration. After 12 months since the 
last  calibration  (Max.  calibration  period)  the  transmitter  will  transmit  this  fact  to  the  host  system.  The  transmitter  must  be  calibrated 
immediately  at  a  certified  service  center  with  a  valid  certificate  of  competance  or  the  manufacturer.  See  section  "Monitoring  the 
calibration peridos". 
 

Technical data and information 

Supply voltage:  

 

 

 

24 V nominal, operational range 8-30 V

 

Cable connections via 4-20 mA:  

 

shielded cable 3 x 1 mm (max. 1200 m) 

 

 

 

 

 

 

shielded cable 3 x 1,5 mm (max. 2400 m) 

Cable connections via RS485: 

 shielded cable 4 x 0,8 mm (max. 400 m) - see section “ instalation of cabeling  
for RS485“ 

 

 

 

 

 

Output:  

 

 

 

 

4 - 20 mA 
RS485 - protocol DEGA, or MODBUS 

 
Standarts: 

 

 

 

 

EN60079-0:2012, IEC60079-0:2011 

 

 

 

 

 

 

EN60079-1:2014, IEC60079-1:2014 

Marking according to ATEX: 

 

 

II 2 G Ex db IIB+H2 T6 Gb Tamb:-40°C-+60°C 

Marking according to IECEx: 

 

 

Ex db IIB+H2 T6 Gb Tamb:-40°C-+60°C

 

II 

– eqiupment group - non-mining environment 

2 G 

- eqiupment  category - explosive atmospheres - zone 1 

Ex d 

- type of protection - flameproof enclosure „d“

 

IIB+H2 

- gas group

 

T6 

– temperature class   

Gb 

- eqiupment  protection level EPL 

Tamb 

– ambient temperature 

 
Dimensions without cable glands: 

 

150 x 170 x 65 mm (WxHxD) 

Weight: 

 

 

 

 

2,1 kg 

Capacity of the internal memory of history: 

44 days at 60 s recording interval 

Interval record storage memory:  

 

60 s (adjustable range 1-255 s) 

Dead band 

 

 

 

 

5 % of range 

 
 
 

 
 
 

Содержание 40100002

Страница 1: ...any form without the prior permission of DEGA CZ s r o is prohibited DEGA CZ s r o reserves the right to alter the specifications of the hardware and software described in this manual at any time and without prior notice DEGA CZ s r o bears no liable for any damage resulting from use of this device ...

Страница 2: ... Additional Ex d cable gland 10 Basic types of transmitters 10 1 Transmitters with a catalytic sensor NSx CL III 10 2 Transmitters with an electrochemical sensor NSx EL III 10 3 Transmitters with an infrared sensor NSx IL III 12 4 Transmitters with a semiconductor sensor NSx SL III 13 5 Transmitters with a PID sensor NSx PL III PID 13 Add on modules 14 Attachments 14 1 Chart for setting the transm...

Страница 3: ...g the functionality of the transmitter and OPERATIONAL AND FUNCTIONAL CHECKS of the entire detenction system sensor excitation with subsequent control of optical and audible alarms triggering fans shutdown technology etc Perform calibration and operational and functional checks only at certified service centers with a valid certificate of competence or the manufacturer Warning The transmitter auto...

Страница 4: ... generation product with new DEGA design DEGA NSx EL III with an electrochemical sensor They operate on the principle of change of electrical parameters on the electrodes stored in electrolyte due to oxidation reduction reactions of the detected gas on it s surface These sensors have good selectivity and the ability to detect very low concentrations of toxic gases DEGA NSx CL III with a catalytic ...

Страница 5: ...e worker exposure to volatile organic compounds VOCs such as solvents fuels degreasers plastics their precursors heat transfer fluids lubricants etc during manufacturing processes and waste handling Product desciption Sensor cover Ex d cable gland OptionalEx d cable gland Body of the removable sensor Sensor bottom Body of the removable sensor Ex d cable gland Mounting holes PCB electronics Ex d ca...

Страница 6: ...ng downwards as specified by the detected substance d Direct the cable trough the cable gland e Connect the cabeling to the terminal block When using the second Ex d cable gland dismout the Ex d blind plug It is allowed to use only Ex d cable glands with M20x1 5 threads f Connect the LCD g Assemble the transmitter with the four hexagon sockets scews 5 mm h Groung the enclosure by a CUPAL washer a ...

Страница 7: ...have a unique adress within the entire bus otherwise there will be communication collisions and malfunctions The transmitter adress can either be set internally using the DegaConfig program or using the DIP switch on the PCB If the position 6 labeled ADD is in the ON position then the adress according to the setting of pins 1 5 is considered Otherwise the adress set in DegaConfig is considered The...

Страница 8: ...tter will transmit the information about the ending calibration via current loop The transmission will be the following 10 s transmitting a 4 20 mA signal informing about the actual gas concentration following a 1 second interval of 2 mA current 5 Reading the record of measured concentrations and alarms The transmitter periodically after 60 s stores the current detected concentration into it s int...

Страница 9: ...recommend to have the air in the deployment area of the detection system analyzed 3 Operation Maintenance In case of contamination the surface can be cleaned with a slightly moistened cloth The seonsors have a different lifetime depending on the sensing technology used and environmental conditions Characteristics of the sensors vary over time Therefore it is required to perform regular checks and ...

Страница 10: ...smitter type Detected gas Measurement range Measurement of current loop 4 20 mA Resolution Calibration gas 40100022 DEGA NSCO EL 1000 III Carbon Monoxide CO 0 1000 ppm 0 130 ppm 1 ppm Carbon Monoxide 450 ppm 40100019 DEGA NSCO EL 200 III Carbon Monoxide CO 0 200 ppm 0 200 ppm 0 1 ppm Carbon Monoxide 130 ppm 40100020 DEGA NBCO EL A500 III Carbon Monoxide CO 0 500 ppm 0 500 ppm 0 1 ppm Carbon Monoxi...

Страница 11: ...III Nitric oxide NO 0 250 ppm 0 15 ppm 0 1 ppm Nitric oxide 15 ppm 40100045 DEGA NSNO EL 1000 III Nitric oxide NO 0 25 ppm 0 25 ppm 0 1 ppm Nitric oxide 150 ppm 40100047 DEGA NSSO2 EL 20 III Nitric oxide NO 0 1000 ppm 0 100 ppm 1 ppm Sulfur dioxide 10 ppm 40100064 DEGA NSSO2 EL 200 III Sulfur dioxide SO2 0 20 ppm 0 3 5 ppm 0 01 ppm Sulfur dioxide 50 ppm 40100065 DEGA NSSO2 EL 2000 III Sulfur dioxi...

Страница 12: ...III Hydrogen fluoride HF 0 10 ppm 0 10 ppm 0 01 ppm VOC 20ppm 40100067 DEGA NSBR2 EL 20 III VOC 0 20 ppm 0 20 ppm 0 01 ppm Chlorine 5 ppm 40100003 DEGA NSBR2 EL 200 III Bromine Br 0 20 ppm 0 20 ppm 0 01 ppm Chlorine 50 ppm 40100004 DEGA NSALC EL 200 III Bromine Br 0 200 ppm 0 200 ppm 0 1 ppm Ethanol 200 ppm 40100002 DEGA NSRCOOH EL 100 III Alkohol 0 200 ppm 0 200 ppm 0 1 ppm Ammonia 300 ppm 401000...

Страница 13: ...0 100 LEL 0 20 LEL 0 1 20 LEL 40100073 DEGA NSR SL 2000 III Refrigerant R401A R404A R407C R32 R410A R12 R22 0 2000 ppm 0 2000 ppm 1 ppm According to the type of refrigerant 40100072 DEGA NSHFO SL 2000 III Refrigerant R1234yf 0 2000 ppm 0 2000 ppm 1 ppm According to the type of refrigerant 5 Transmitters with a PID sensor NSx PL III PID Product code Transmitter type Detected gas Measurement range M...

Страница 14: ... OFF ON ON OFF ON 23 ON ON ON OFF ON 24 OFF OFF OFF ON ON 25 ON OFF OFF ON ON 26 OFF ON OFF ON ON 27 ON ON OFF ON ON 28 OFF OFF ON ON ON 29 ON OFF ON ON ON 30 OFF ON ON ON ON 31 ON ON ON ON ON 2 Signalization transmitted by the current loop 4 20 mA Measurement The measure concentration is directly proportional to 4 20 mA current output Exceeding the range of measured concentrations Current output ...

Страница 15: ...ct shall be replaced 616 paragraph 4 of the Commercial Code 6 The buyer who exercises the right of warranty repair is not entitled to the return of the parts that have been replaced 7 The warranty period can be extended for up to 48 months and its validity can be extended beyond the standard length on the basis of the conclusion of an individual warranty contract Further information may be obtaine...

Отзывы: