Management Procedure 2559 Rev A
Page
6
of
7
4.1.6
Set the UUC to 67.5V, 90 k
Ω
and adjust the decade box to 90 k
Ω
. Record the voltage
displayed on the multimeter.
4.1.7
Set the UUC to 90V and adjust the decade box to 400 k
Ω
. Record the voltage displayed
on the multimeter.
5
Performance Requirements
Table 5-1 Performance Requirements and Calibration Data for PosiTest LPD
Parameter
Nominal
Min
1
UUC
Reading Max
2
Resistance
90 k
Ω
(9V)
87.3
92.7
Resistance
80 k
Ω
(67.5V)
77.6
82.4
Resistance
90 k
Ω
(67.5V)
87.3
92.7
Resistance
400 k
Ω
(90V)
388
412
DC Voltage
9V (90 k
Ω
)
8.55
9.45
DC Voltage
67.5V (80 k
Ω
)
64.13
70.87
DC Voltage
67.5V (90 k
Ω
)
64.13
70.87
DC Voltage
90V (400 k
Ω
)
85.50
94.50
1)
Resistance min = nominal value (k
Ω
) * 0.97
DC Voltage min= nominal value (V) * 0.95
2)
Resistance max = nominal value (k
Ω
) * 1.03
DC Voltage max= nominal value (V) * 1.05
Note
: Do not write in this procedure.