Page 10
Calibration adjustments are returned to the Gage’s factory settings. The factory
calibration icon will appear on the display.
Menu settings are returned to the following:
Memory = OFF
Auto Sub-Batch = OFF
Statistics Mode = OFF
HiLo Alarm = OFF
Bluetooth & Stream = OFF
BLE Keyboard = OFF
WiFi & Access Point = OFF
USB Keyboard & Stream = OFF
Auto Dim = ON
Hi Res = OFF
Perform a more thorough Hard Reset by powering down the Gage, waiting several
seconds, then simultaneously holding both the center navigation and (+) buttons until
the Reset symbol appears. This returns the instrument to a known, out-of-the-box
condition. It performs the same function as a menu Reset with the addition of:
Bluetooth Pairing info is cleared. Menu settings are returned to the following
Bluetooth Smart = OFF
Test Time = 1 second
Sound = Medium
Touch = ON
Flip Lock = OFF
Auto Sync .net = ON
USB Drive = ON
Language = English
Battery Type = Alkaline
NOTE:
Date, Time and WiFi are not affected by either Reset.
Battery Type
Selects the type of batteries used in the Gage from a choice of Alkaline, Lithium or
NiMH (nickel-metal hydride rechargeable). The battery state indicator symbol is
calibrated for the selected battery type. No damage will occur if the battery type used in
the Gage does not match the selected battery type.