8
Auto Log Mode
Auto Log Mode
The
PosiTector
DPM
can automatically display and record
datasets at user selected time intervals. When selected, the Auto
Log icon will appear on the display, with a countdown timer until
the next reading. Connect the instrument to a USB power source
for long-term use.
Turn on instrument/
Cancel Auto Log
mode
Scroll through
display modes
(Advanced only)
Standard
models
remain powered-up while in auto log mode. The
instrument will take readings for 50-60 hours on a set of alkaline
batteries, depending on the frequency of measurements.
Advanced
models
power down between readings to conserve
batteries. At each interval, the instrument will wake up, take a
reading, sync to PosiSoft.net (if configured) and power down. The
instrument can record approximately 10,000 datasets while in this
mode using alkaline batteries, and can be configured to
Auto
Batch
at daily, weekly, and monthly intervals.
-
Auto Log
continues recording until the instrument loses power or
the user turns off
Auto Log
mode. Stored datasets will remain in
memory until full, then the oldest dataset or batch will be overwrit-
ten to allow space for the next dataset.
-To conserve battery life, readings update only at the Log Interval.
The last recorded readings will remain on the display until the
next dataset has been stored.
-It is recommended that fresh alkaline batteries be installed prior
to an extended
Auto Log
if not connected to continuous power.