3.2. Ratings
3.2.1. Detector
Table 3.3:
Power Ratings
Detector power input
2 x 100 VAC to 240 VAC, 50 Hz to 60 Hz, 580 W
Power supply 1 (modules): 360 W
Power supply 2 (electronics): 220 W
Can be connected to all common supply voltages.
Detector external trigger input
2
.
1 V to 5
.
0 V high level
0
.
0 V to 0
.
8 V low level
Caution
#3
Absolute maximum is 5 V. Applying a higher voltage
will damage the detector.
External trigger input impedance
50 Ω
Detector trigger output
5 V (max. current 100 mA)
3.2.2. Detector Control Unit
Table 3.4:
Detector Control Unit Ratings
Power input
2 x 100 VAC to 240 VAC, 50
/
60 Hz, 3
.
7 A to 7
.
4 A, 550 W (Plat-
inum)
1+1 redundant, hot swappable power supply unit
Dimensions (W x H x D)
482
.
4 mm (434
.
0 mm without rack latches) x 42
.
8 mm x 715 mm
Weight
16
.
73 kg to 19
.
9 kg
Chassis
1U
3.2.3. Thermal Stabilization Unit
Table 3.5:
Thermal Stabilization Unit Ratings
Power input
Single phase 100 VAC to 240 VAC, allowable voltage range
±
10 %, 50/60 Hz, 8 A (100 VAC) to 3 A (240 VAC)
Dimensions (W x H x D)
270 mm x 393 mm x 436 mm
Weight
17
.
5 kg
Typical flow
3 L min
−
1
Maximum operation pressure
3 bar
PILATUS3 X 6M Technical Specifications v1.2.0
6 | 26