4.1.4. Vacuum Feedthroughs (Optional)
For vacuum applications separate electrical and cooling feedthrough sets are optionally available.
The mechanical interface of the electrical feedthrough is a standard DN100ISO-K flange and for the cooling feedthrough it is
a standard DN40ISO-KF flange. In the standard configuration the cable length is 1 m in-vacuum and 1 m in-air.Their lengths
can be adjusted between 0.5 and 1 m. (The functionality of the system with a cable length of >2 m is not guaranteed) The
cooling hoses are convoluted stainless steel tubes, suitable for static applications, but not for dynamic/flexing applications.
Caution
#7
The cable of the feedthrough set should not be combined with the default cables of the detector system.
Figure 4.9:
Picture of the optional PILATUS3 R CdTe 100K-M electrical vacuum feedthrough viewed from the vacuum (left) and air side
(right).
Figure 4.10:
Picture of the optional PILATUS3 R CdTe 100K-M cooling vacuum feedthrough set with tubes through a standard DN40ISO-
KF flange, and straight and elbow connectors for the detector head.