7.2. Error Codes
If the response of the server consists of a single, negative number, an error has occured during the execution of
the command. The following table lists the corresponding error codes.
Table 8 List of error codes.
Error code
Description
-1
Unknown command
-2
Invalid argument
-3
Unknown settings
-4
Out of memory
-5
Module calibration files not found
-6
Readout failed
-10
Flatfield file not found
-11
Bad channel file not found
-12
Energy calibration file not found
-13
Noise file not found
-14
Trimbit file not found
-15
Invalid format of the flatfield file
-16
Invalid format of the bad channel file
-17
Invalid format of the energy calibration file
-18
Invalid format of the noise file
-19
Invalid format of the trimbit file
-20
Version file not found
-21
Invalid format of the version file
-22
Gain calibration file not found
-23
Invalid format of the gain calibration file
-24
Dead time file not found
-25
Invalid format of the dead time file
-30
Could not create log file
-31
Could not close log file
-32
Could not read log file