V S A   Q u i c k   S t a r t   G u i d e

Fi

gure 3: Tank fill extension

  •  Remove the black plug on the clear plastic tank fill

 

    extension located on the back right of the VSA. 

 

    (see Figure 3: Tank fill extension). Fill the tank with

 

    deionized water until full. Reinstall the black plug

 

    once finished. It is not necessary to tighten the lid with a

 

    screwdriver, hand-tight is good enough.

 

  •  To remove the water from the water tank, the water tank

 

    drain plug is used.

  •  Figure 3 shows the location of a filter that will need to be

 

    removed from time to time when dust builds up on it. The

 

    housing around the filter is a snap on housing that is easily

 

    removed. 

Water Fill Plug

Water Tank Fill Extension

Upper Horizontal Drilled Hole

Water Tank Drain Plug

Fan Filter

V S A   Q u i c k   S t a r t   G u i d e

        10-600 ml/min can be selected. Higher flow rates will

 

        speed up the test, but may result in lower data

 

        resolution and non-equilibrium conditions. The

 

        default value of 100 ml/min is suitable for most

 

        products.

     

• 

To choose to include a timeout setting, press   with

 

        “Timeout” highlighted, and then cycle to the desired

 

        value using the up and down arrows. Any value

 

        between 5.0 min to 30 days can be selected. The

 

        timeout determines the maximum time allotted to

 

        move from the starting 

a

w

 to the final 

a

w

 (DDI only)

 

        and is optional. The default value is off, which means

 

        there is no timeout value.

     

• 

To choose to have the isotherm loop, press   with

 

        “Loop” highlighted to toggle between on and off. 

 

        Turning loop on causes the isotherm to automatically

 

        return to the start 

a

w

 (with the same settings) once

 

        the final 

a

w

 has been achieved. This would create

 

        both an adsorption and desorption curve (or vice versa

 

        depending on the starting direction).

  •  For DVS method

     

• 

To choose the starting 

a

w

, press   with “Start”

 

        highlighted, and then cycle to the desired value using

 

        the up and down arrows. Any value between 0.03

 

        and 0.95 

a

w

 can be chosen including “current”. 

 

        Choosing “current” means that the stage will start at

 

        the initial 

a

w

 of the sample. Press   when finished.

     

• 

To choose the final 

a

w

, press   with “Final”

 

        highlighted, and then cycle to the desired value using

 

        the up and down arrows. Any value between 0.03

 

        and 0.95 

a

w

 can be chosen including current 

 

        (see“current” definition above). Selecting a final 

a

w

 

        that is lower than the starting 

a

w

 will result in

 

        desorption, while selecting a final 

a

w

 higher than the

 

        starting 

a

w

 will result in adsorption. Press   when

 

        finished.

     

• 

To choose the temperature of the stage, press   with

 

        “Temperature” highlighted, and then cycle to the

 