V S A   Q u i c k   S t a r t   G u i d e

  •  For DDI method

  

• 

To choose the starting a

w

, press   with “Start” 

      highlighted and then cycle to the desired value using 

      the up and down arrows. Any value between 0.03 and 

           0.95 a

w

 can be chosen including “current”.  Choosing 

           “current” means that the stage will start at the initial a

w

 

           of the sample. Press   when finished.

    •  To choose the final a

w

, press   with “Final” highlighted 

           and then cycle to the desired value using the up and 

           down arrows. Any value between 0.03 and 0.95 a

w

 can 

           be chosen including current (see “current” definition 

           above). Selecting a final a

w

 that is lower than the 

           starting a

w

 will result in desorption, while selecting a 

           final a

w

 higher than the starting a

w

 results in adsorption. 

           Press   when finished.

  

• 

To choose the temperature of the stage, press   with 

      “Temperature” highlighted and then cycle to the 

      desired temperature using the up and down arrow keys 

      (hold down arrows for accelerated scrolling). You may 

           choose any temperature between 15 and 60 °C, 25 °C is 

           default. Each stage can run at a unique temperature.

  

• 

To choose the desired 

a

w

 resolution, highlight 

           “Resolution” press   with “Resolution” highlighted and 

           then cycle to the desired value using the up and down 

           arrows. You may choose any value between 0.003 

           and 1.0 

a

w

. The resolution for DDI determines the target 

           

a

w

 resolution. Setting a higher resolution value makes 

           the test faster, but decreases the 

a

w

 resolution. We 

           recommend a DDI resolution of 0.01 

a

w

.  

  

• 

To choose the air flow rate, press   with “Flow” 

      highlighted, and then cycle to the desired value using 

      the up and down arrows. Any value between  

      10-163 ml/min can be selected. Higher flow rates will 

      speed up the test, but may result in lower data 

      resolution and non-equilibrium conditions. The 

      default value of 100 ml/min is suitable for most 

      products.

Содержание AQUA LAB VSA

Страница 1: ...Vapor Sorption Analyzer Quick Start Guide...

Страница 2: ...nmental temperature and humidity remain fairly stable Figure 1 Adjustment and bubble level The AquaLab VSA needs to be on a stable platform and evaluated to verify shipping did not damage the scale pe...

Страница 3: ...esiccant installation Desiccant tubes are installed by orienting the tube so that the black colored valve inserts into the black receiving hole as shown in Figure 2 Before insertion make sure that the...

Страница 4: ...er tighten the water fill plug once finished To remove the water from the water tank the water tank drain plug is used Figure 3 shows the location of a filter that will need to be removed from time to...

Страница 5: ...cup For best results verify cleanliness of all parts 2 AquaLab VSA Performance Verification overview AquaLab VSA performance and cleanliness should be checked frequently before each isotherm test is...

Страница 6: ...The aw reading for the standard and the temperature displays when finished Make sure the aw reading on the standard is within 0 005 of the correct aw for the standard 0 02 if cap sensor is used If an...

Страница 7: ...e settings for that portion of the isotherm Once a stage is created it can be modified or deleted anytime before the test is started i e For milk powder Stage 1 could use the DDI method to ramp quickl...

Страница 8: ...e highlighted and then cycle to the desired temperature using the up and down arrow keys hold down arrows for accelerated scrolling You may choose any temperature between 15 and 60 C 25 C is default E...

Страница 9: ...art highlighted and then cycle to the desired value using the up and down arrows Any value between 0 03 and 0 95 aw can be chosen including current Choosing current means that the stage will start at...

Страница 10: ...to indicate equilibrium for a given step Setting a higher value will make the test faster but may not result in complete equilibrium If the trigger is turned off the sample will be held at each step...

Страница 11: ...be done when isotherm test is running Remove the weighing pan Press until you are at the Measurement Tab 2nd tab Move latch left and or press to start a reading The pinwheel will spin indicating a me...

Страница 12: ...2365 NE Hopkins Court Pullman Washington 99163 USA 1 509 332 2756 instruments aqualab com www aqualab com 14510 13 01 16...

Отзывы: