Vapor Sorption 

Analyzer

Quick Start Guide

Содержание AQUA LAB VSA

Страница 1: ...Vapor Sorption Analyzer Quick Start Guide...

Страница 2: ...nmental temperature and humidity remain fairly stable Figure 1 Adjustment and bubble level The AquaLab VSA needs to be on a stable platform and evaluated to verify shipping did not damage the scale pe...

Страница 3: ...esiccant installation Desiccant tubes are installed by orienting the tube so that the black colored valve inserts into the black receiving hole as shown in Figure 2 Before insertion make sure that the...

Страница 4: ...er tighten the water fill plug once finished To remove the water from the water tank the water tank drain plug is used Figure 3 shows the location of a filter that will need to be removed from time to...

Страница 5: ...cup For best results verify cleanliness of all parts 2 AquaLab VSA Performance Verification overview AquaLab VSA performance and cleanliness should be checked frequently before each isotherm test is...

Страница 6: ...The aw reading for the standard and the temperature displays when finished Make sure the aw reading on the standard is within 0 005 of the correct aw for the standard 0 02 if cap sensor is used If an...

Страница 7: ...e settings for that portion of the isotherm Once a stage is created it can be modified or deleted anytime before the test is started i e For milk powder Stage 1 could use the DDI method to ramp quickl...

Страница 8: ...e highlighted and then cycle to the desired temperature using the up and down arrow keys hold down arrows for accelerated scrolling You may choose any temperature between 15 and 60 C 25 C is default E...

Страница 9: ...art highlighted and then cycle to the desired value using the up and down arrows Any value between 0 03 and 0 95 aw can be chosen including current Choosing current means that the stage will start at...

Страница 10: ...to indicate equilibrium for a given step Setting a higher value will make the test faster but may not result in complete equilibrium If the trigger is turned off the sample will be held at each step...

Страница 11: ...be done when isotherm test is running Remove the weighing pan Press until you are at the Measurement Tab 2nd tab Move latch left and or press to start a reading The pinwheel will spin indicating a me...

Страница 12: ...2365 NE Hopkins Court Pullman Washington 99163 USA 1 509 332 2756 instruments aqualab com www aqualab com 14510 13 01 16...