EC-5 User’s Manual
4. Collecting Data
11
01: 1 Execution Interval (seconds)
; Factory calibration equations for ECH2O
; probes convert mV output of ECH2O to
; volumetric water content (VWC, m3/m3)
; EC-5: VWC = 0.00119 * mV - 0.400
1: Excite-Delay (SE) (P4)
1: 1 Reps
2: 5 2500 mV Slow Range
3: 1 SE Channel
4: 1 Excite all reps w/Exchan 1
5: 1 Delay (0.01 sec units)
6: 2500 mV Excitation
7: 1 Loc [ Probe_VWC ]
8: .00119 Multiplier
9: -.4 Offset
*Table 2 Program
02: 0.0000 Execution Interval (seconds)
*Table 3 Subroutines
End Program
SCWin (Short Cut) Directions
The following are instructions for using Campbell Sci-
entific’s SCWin (Short Cut) program to read the EC-5
soil moisture sensor.
1. Download EchoCSI.zip from http://www.deca-
gon.com/appnotes/EchoCSIappnote.pdf.
2. Unzip the folder EchoCSI.zip.
3. Locate the file containing SCWin.exe. It should be
in C:\Program Files\Campbellsci\SCWin. Place the
Содержание EC-5