background image

     

 

  1 / 17 

 

WWW.DDTOP.COM 

 

 

UHC | Bb-Remote Transmission   
 
MAGNETIC LEVEL GAUGE 

 

 

Operating Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U

HC

-Bb
-DT
-JS
-10
2

7

-20
18
(A

Содержание UHC Magnetic Level Gauge

Страница 1: ...1 17 WWW DDTOP COM UHC Bb Remote Transmission MAGNETIC LEVEL GAUGE Operating Manual UHC Bb DT JS 1027 2018 A...

Страница 2: ...and are subject to change without notice Dandong Top Electronics Instrument Group Co Ltd reserves the right to modify the products described in this manual at any time without prior notice Quality As...

Страница 3: ...arranty Quality Dandong Top Electronics Instrument Group Co Ltd has passed the ISO9001 quality system certification The whole process of product production is strictly in accordance with the scope of...

Страница 4: ...Product Descriptions 5 3 Technical Parameters 6 4 Outline Dimension Diagram 7 5 Unpacking and Inspection 8 6 Installation 8 7 Configuration 10 8 Precautions 15 9 Fault Analysis and Troubleshooting 16...

Страница 5: ...onal standards codes and regulations Ensure that the equipment is installed in accordance with intrinsically safe or non flammable site operating procedures 1 2 Process leakage may cause serious injur...

Страница 6: ...y or output to the control room through the output terminal to achieve long distance transmission of liquid level 2 3 Packaging Please take the packaging waste to a special recycling facility 2 4 Lift...

Страница 7: ...40 80 Working Temperature 40 200 When T 150 the transmitter does not allow heat preservation Accuracy 5mm Power Inlet M20 1 5 female thread or as customer requirement Explosion proof Type Intrinsicall...

Страница 8: ...product nameplate is consistent with the packing list information Figure 4 Figure 4 Product Nameplate 5 1 2 Check the quantity and material of each part against the packing list 5 2 Content of Inspect...

Страница 9: ...point full scale of the remote transmission device marked at the factory and the O scale full scale of the on site indicator on the measuring tube are one to one correspondence and then fixed with a...

Страница 10: ...lay a single variable or you can set the percentage and the main variable to cycle display and the effective digits of the decimal part can be set 0 1 2 3 4 decimal places The display mode is set by 4...

Страница 11: ...n the 888 character in the lower left corner shows 1 17 it indicates that the transmitter is in the field configuration mode with menu rotation At this time you can input the initialization code modif...

Страница 12: ...the key the up and down arrows on the left will light up indicating that the setting is saved 7 Press again to restart setting data 7 1 5 Key function code quick reference table When using button conf...

Страница 13: ...n arrows on the left will light up Unit Setup 7 1 7 Range setting Set the lower limit of the range LRV Long press the key to enter the configuration mode first menu No 1 is displayed the up and down a...

Страница 14: ...ation First adjust the PV value to the zero position Move the magnetic float to zero Switch the menu to No 11 and the LCD will display the current calibration position value at the same time and then...

Страница 15: ...confirm and start the next digit setting and set other digits and decimal points in sequence When the up and down arrows on the left side light up the menu function setting is complete Compensation v...

Страница 16: ...n the container etc Please refer to the instructions in the chapter 6 2 Installation Operation Process and 7 1 Electrical Wiring and remove the components in the reverse order of the operation steps 1...

Страница 17: ...17 17 WWW DDTOP COM...

Отзывы: