Quick
S
et-up operation
1. Determine the Peak Mode required
Configure switch settings 2 and 3 for the proper function.
2. Select FAST or SELECT signal source
Configure switch 1 - FAST will be the highest dynamic capture. SELECT
will reflect the low pass filter selection on the main 3700 Analog Controls.
3. Select the Front Panel Operaton of the TARE Switch
Configure switch
5
- Normal operation of the TARE switch
to provide "zero-ing"
of the analog signal
or
to use the TARE switch
as a
front panel
PEAK -TRACK
"push-push" operation to
place the
3700
unit is PEAK capture Mode or Reset the
PEAK that has been held
resulting in the display and analog output signal tracking
the live input
.
Push IN - PEAK. Push OUT - TRACK.
4. Select the Operation of the Display
Switch 4 -
Meter display can be configured for LIVE display or the PEAK capture
value. If the unit is a "stand-alone" meter with the
local
operator providing the
Peak
Reset
,
as mentioned in step 3, the
n Switch 4 should be DOWN for
PEAK
display
operation. Remote control of the display can also be done via the I/O
connections with an external switch or controller providing the Display control
function.
5. Determine how much "threshold" level is needed for your application
Switch 6 will display the threshold value (
1
to
8
%) for a bi-polar "no peak
detection zone" around
the
Zero measurement level. This adjustment is normally
in
the 5% area, but can be adjusted depending on the application and how much
PEAK defeat /
rejection is needed. When finished with
the
potentiometer
adjustment, return
Switch 6 to the UP position.
6
. Determine
the need for remote control of
your application
Refer to the Connector I/O interface in section 3 for
setup functions
of
the 8
positon
termi
n
al connector.
Switch function 4 and 5 can be
remotely controlled. All other switch settings are typically configured
during inital setup of the 3700 unit and will not be changed.
Page
9