DS-1000 Operating Manual
Maintenance
page 25
carry-on baggage A shipping container with foam packing is supplied as
standard equipment with all DS-1000 curve tracers. Daystar strongly
recommends using this case anytime the DS-1000 is shipped.
DS-Tracer Fuses
The DS-Tracer has two externally accessible fuses. The first of the external
fuses is located in the AC input power receptacle. Power line surges or internal
faults could cause this fuse to fail. Return the DS-Tracer for repairs if this fuse
blows repeatedly. Replace this fuse with the type indicated on the DS-Tracer
panel.
A second user replaceable fuse is the 2 Amp, fast-blow DC charging fuse. This
fuse should only blow if a DC source voltage exceeds the rating on the DS-
Tracer panel. Replace only with a 2 Amp, fast-blow DC charging fuse.
There is an internal fuse on the power supply board battery circuit. This fuse is
installed as protection when the DS-Tracer is removed from its case during
repairs.