![DAYLIGHT SOLUTIONS MIRcat-QT 2 Series Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/daylight-solutions/mircat-qt-2-series/mircat-qt-2-series_user-manual_3890228009.webp)
MIRcat OPERATION
•
D11-00028-02-A
Page
9
of
54
5.3 Cooling Lines – Preventive Maintenance
Periodically, and after moving the laser system, inspect the system for contamination and verify that
cooling ports are clear and all cable connections are visibly secure. If liquid cooling is employed, verify the
cooling connections are secure. For cleaning and decontamination of the laser system, use only mild
detergent and moist paper towel. Do not submerse in water or cleaning solvent.
5.4 Laser Safety Preventive Maintenance – Maintaining Regulatory
Compliance.
It is recommended that all laser safety features and safety interlocks included in your system are tested and
their proper function verified every 6 months to ensure they are still operating correctly. If any safety
feature or interlock fails to perform as described in Section VI of this manual, stop use of the laser system
immediately, prevent others from using it (by removing the key, for example), and contact Daylight
Solutions or your authorized representative for further instructions.